Instruction Manual

pH 1100/ 2100

54

10.5.8

Memory clear setup mode P4.6


Clears all pH, mV and Ion (pH2100 only) memory points stored
in memory.

1. Enter meter setup mode as described in steps 1-4 in Section

10.5.1 Entering meter setup mode (page 50).


2. Use

/MI key to scroll to program Mem Clear Setup: P4.6.

See Figure 181.


3. Press

ENTER to enter program Mem Clear Setup: P4.6.


4. Press

>/CAL key to toggle between Memory Clear Yes and No.

The selection will flash. See Figure 182.


5. Press

ENTER to confirm selection.


• If you select Yes, the screen flashes Memory being

Cleared, Please wait... See Figure 183. The meter then
returns to the setup mode.

• If you select No, the meter returns to the setup mode.

Press

/MI to continue to the next setup program, or

press

/MR to go to the previous program. Press EXIT to

return to measurement mode.




10.5.9

Meter reset setup mode P4.7


Resets the meter to factory default settings:
• pH/mV/Ion* calibration data is erased
• Manual temperature compensation is set to 25.0°C
• pH/mV/Ion* measurement alarm is off
• pH/mV/Ion* calibration due alarm is off
• Stability indicator is turned on with low stability setting
• Data logging and memory recall has manual setting
• Backlight is on with 2.0 minute duration
• Memory is cleared
*Ion features on the CyberScan pH2100 meter only

1. Enter meter setup mode as described in steps 1-4 in Section

10.5.1 Entering meter setup mode (page 50).


2. Use

/MI key to scroll to program Meter Reset: P4.7.

See Figure 184.


3. Press

ENTER to enter program Meter Reset: P4.7.


4. Press

>/CAL key to toggle between Meter Reset Yes and No.

The selection will flash. See Figure 185.


5. Press

ENTER to confirm selection.

• If you selected Yes, the screen flashes Meter being

Reset, Please wait...See Figure 186. The meter then
returns to measurement mode.

• If you select No, the screen flashes Reset Aborted. See

Figure 187. The meter then returns to the measurement
mode.

Memory being Cleared
Please wait. . .

Memory Clear:Yes

No

Select & Press ENTER

Mem Clear Setup :P4.6

Press ENT. to Select

Figure 181

Figure 182

Figure 183

Meter being Reset

Please wait. . .

Meter Reset: Yes

No

Select & Press ENTER

Meter Reset :P4.7

Press ENT. to Select

Figure 184

Figure 185

Figure 186

Figure 187

Reset Aborted