10.5.8
Memory clear setup mode P4.6
Clears all pH, mV and Ion (pH2100 only) memory points stored
in memory.
1. Enter meter setup mode as described in steps 1-4 in Section
10.5.1 Entering meter setup mode (page 50).
2. Use
∆/MI key to scroll to program Mem Clear Setup: P4.6.
See Figure 181.
3. Press
ENTER to enter program Mem Clear Setup: P4.6.
4. Press
>/CAL key to toggle between Memory Clear Yes and No.
The selection will flash. See Figure 182.
5. Press
ENTER to confirm selection.
• If you select Yes, the screen flashes Memory being
Cleared, Please wait... See Figure 183. The meter then
returns to the setup mode.
• If you select No, the meter returns to the setup mode.
Press
∆/MI to continue to the next setup program, or
press
∇/MR to go to the previous program. Press EXIT to
return to measurement mode.
10.5.9
Meter reset setup mode P4.7
Resets the meter to factory default settings:
• pH/mV/Ion* calibration data is erased
• Manual temperature compensation is set to 25.0°C
• pH/mV/Ion* measurement alarm is off
• pH/mV/Ion* calibration due alarm is off
• Stability indicator is turned on with low stability setting
• Data logging and memory recall has manual setting
• Backlight is on with 2.0 minute duration
• Memory is cleared
*Ion features on the CyberScan pH2100 meter only
1. Enter meter setup mode as described in steps 1-4 in Section
10.5.1 Entering meter setup mode (page 50).
2. Use
∆/MI key to scroll to program Meter Reset: P4.7.
See Figure 184.
3. Press
ENTER to enter program Meter Reset: P4.7.
4. Press
>/CAL key to toggle between Meter Reset Yes and No.
The selection will flash. See Figure 185.
5. Press
ENTER to confirm selection.
• If you selected Yes, the screen flashes Meter being
Reset, Please wait...See Figure 186. The meter then
returns to measurement mode.
• If you select No, the screen flashes Reset Aborted. See
Figure 187. The meter then returns to the measurement
mode.
Memory being Cleared
Please wait. . .
Memory Clear:Yes
No
Select & Press ENTER
Mem Clear Setup :P4.6
Press ENT. to Select
Figure 181
Figure 182
Figure 183
Meter being Reset
Please wait. . .
Meter Reset: Yes
No
Select & Press ENTER
Meter Reset :P4.7
Press ENT. to Select
Figure 184
Figure 185
Figure 186
Figure 187
Reset Aborted