10.5 Meter general configuration setup mode
10.5.1
Entering meter general setup mode
1. Turn the meter ON. The meter setup mode can be entered
from any measurement mode—pH, mV or ion (pH 2100 meter
only).
2. Press
SETUP to enter the setup mode. The display will flash
Setup Mode Activated, Please wait. ..See Figure 159.
3. Use
>/CAL to select "Meter". The mode selected will flash.
See Figure 160, 161 and 162. Press ENTER to enter Meter
setup mode. You are now in the Meter setup mode.
4. The first setup program in Ion mode is Date/Time Setup: P4.0.
See Figure 163. If you want to skip to another program, use
the
∆/MI and ∇/MR keys to scroll to the desired program.
5. To exit the setup mode at anytime, press EXIT. The meter
returns to the measurement mode that it was in prior to
entering meter setup mode.
NOTE: To exit the setup mode at anytime without confirming
changes, press EXIT. If you have made any changes to any
programs and press EXIT, these changes will be confirmed. To
reset meter to default factory settings, see setup program Meter
Reset P4.7 (page 54).
10.5.2
Date/time setup mode P4.0
Use this program to set the meter date and time.
1. Enter the Meter setup mode as described above in Section
10.5.1 Entering meter general setup mode (Pg 50). The first
setup mode is Date/Time Setup: P4.0. See Figure 163.
2. Press
ENTER to enter program Date/Time Setup: P4.0.
3. Key in the Date, Month, Year, Hr and Min using the numeric
keypad. Use the >/CAL key or the ENTER key to toggle to each
field. If you make a mistake keying the value, use the
∆/MI and
∇/MR keys to highlight mistake, then re-enter.
See Figure 164.
4. After keying in a value for the last field (Min), press ENTER to
confirm. The screen flashes Clock Updating, Please wait. . See
Figure 165. The clock then flashes the date and time.
See Figure 166.
Press
∆/MI key to continue to the next setup program, or press
∇/MR to go to the previous program. Press EXIT to return to the
measurement mode you were in prior to entering meter setup
mode.
Setup: mV Meter
Select & Press ENTER
Setup:pH/Temp Meter
Select & Press ENTER
Setup Mode Activated
Please wait. . .
Figure 159
Figure 160
Figure 161
Figure 162
Setup: Ion Meter
Select & Press ENTER
Clock Updating
Please wait. . .
Date:23 Month:01
Year:02 Hr:8 Min:07
Date/Time Setup:P4.0
Press ENT. to Select
Figure 163
Figure 164
Figure 165
Figure 166
Date Time
Jan 23 2002 08”07:31