Reset aborted ph being reset please wait, Ph reset: yes no select & press enter, Ph cal reset :p1.7 press ent. to select – Thermo Fisher Scientific CyberScan pH 1100/pH 2100 User Manual
Page 36
![background image](/manuals/370643/36/background.png)
Instruction Manual
pH 1100/ 2100
36
10.2.9
pH calibration data reset setup program P1.7
Use this program to erase all pH calibration data.
1. Enter pH/Temperature setup mode as described in steps 1-4 in
Section 10.2.1 Entering pH/temperature setup mode (page 30).
2. Press
∆/MI to scroll to program pH Cal Reset: P1.7. See
Figure 105.
3. Press
ENTER to enter program pH Cal Reset: P1.7.
4. Press
>/CAL to toggle between pH Reset: Yes and No. The
selection will flash. See Figure 106.
5. Press
ENTER to confirm selection.
• If you selected Yes, the screen flashes pH being Reset,
Please wait... See Figure 107. The meter then returns to
setup mode.
• If you select No, the screen flashes Reset Aborted and
then returns to the setup mode. See Figure 108.
This is the last pH setup program. Press
∇/MR to go to the previous
program. Press
∆/MI to go back to the first pH setup program. Press
EXIT to return to the pH measurement mode.
Reset Aborted
pH being Reset
Please wait. . .
pH Reset: Yes No
Select & Press ENTER
pH Cal Reset :P1.7
Press ENT. to Select
Figure 105
Figure 106
Figure 107
Figure 108