Degree c degree f select & press enter, Temp unit setup:p1.2 press ent. to select, Ph resol. setup: p1.1 press ent. to select – Thermo Fisher Scientific CyberScan pH 1100/pH 2100 User Manual
Page 31
![background image](/manuals/370643/31/background.png)
Instruction Manual
pH 1100/ 2100
31
10.2.3
pH resolution setup program P1.1
Use this program to select pH resolution (0.01 or 0.001 pH).
1. Enter
pH/Temperature
setup
mode as described in steps 1-4
in 10.2.1 Entering pH/temperature setup mode (page 30).
2. Use
∆/MI key to scroll to program pH Resol. Setup: P1.1. See
Figure 83.
3. Press
ENTER to enter program pH Resol. Setup: P1.1.
4. Press >/CAL to toggle between 0.01 and 0.001. The
resolution selected will flash. See Figure 84.
5. Press
ENTER to confirm selection and return to the setup
mode. Press
∆/MI to continue to the next setup program, or
press
∇/MR to go to the previous program. Press EXIT to
return to the pH measurement mode.
10.2.4
Temperature unit setup program P1.2
Use this program to select temperature measurement unit (°C or
°F).
1. Enter pH/Temperature setup mode as described in steps 1-4
in 10.2.1 Entering pH/temperature setup mode (page 30).
2. Use
∆/MI key to scroll to program Temp Unit Setup: P1.2. See
Figure 85.
3. Press
ENTER to enter program Temp Unit Setup: P1.2.
4. Press
>/CAL to toggle between Degree C and Degree F. The
unit selected will flash. See Figure 86.
5. Press
ENTER to confirm selection and return to the setup
mode. Press
∆/MI to continue to the next setup program, or
press
∇/MR to go to the previous program. Press EXIT to
return to the pH measurement mode.
Mode
Degree C Degree F
Select & Press ENTER
0.01 0.001
Select & Press ENTER
Temp Unit Setup:P1.2
Press ENT. to Select
pH Resol. Setup: P1.1
Press ENT. to Select
Figure 83
Figure 84
Figure 85
Figure 86