Thermo Fisher Scientific CyberScan pH 1100/pH 2100 User Manual
Page 37
![background image](/manuals/370643/37/background.png)
Instruction Manual
pH 1100/ 2100
37
10.3 mV setup mode
10.3.1
Entering mV setup mode
1. Turn the meter ON. If the meter is not in the mV measurement
mode, press EXIT and use >/CAL to select mV. The mode
selected will flash. Press ENTER to confirm selection. Their
meter is now in mV measurement mode.
2. Press
SETUP to enter the setup mode. The display will flash
Setup Mode Activated, Please wait... See Figure 109.
3. Press
>/CAL to select mV. The mode selected will flash. See
Figure 110. Press ENTER to enter mV setup mode. You are
now in the mV setup mode.
4. The first setup program in mV mode is mV Meas. Alarm: P2.0.
See Figure 111. If you want to skip to another program, use
the
∆/MI and ∇/MR keys to scroll to the desired program.
5. To exit the setup mode at anytime, press EXIT. The meter
returns to the mV measurement mode.
NOTE: To exit the setup mode at anytime without confirming
changes, press EXIT. If you have made any changes to any
programs and press EXIT, these changes will be confirmed. To
reset meter to default factory settings, see setup program Meter
Reset P4.7 (page 54).
mV Meas. Alarm :P2.0
Press ENT. to Select
Setup Mode Activated
Please wait. . .
Setup: mV
Meter
Select & Press ENTER
Figure 109
Figure 110
Figure 111