Cal alarm due on off select & press enter, Mvcaldue alarm :p2.1 press ent. to select – Thermo Fisher Scientific CyberScan pH 1100/pH 2100 User Manual
Page 39
![background image](/manuals/370643/39/background.png)
Instruction Manual
pH 1100/ 2100
39
10.3.3
mV calibration due alarm setup program P2.1
Use this program to set a reminder when the next calibration is
due. Choose from 1 to 365 days from last calibration.
1. Enter mV setup mode as described in steps 1-4 in Section
10.3.1 Entering mV setup mode (page 37).
2. Use
∆/MI key to scroll to program mVCalDue Alarm: P2.1.
See Figure 116.
3. Press
ENTER to enter program mVCalDue Alarm: P2.1.
4. Press
>/CAL to toggle between Cal Alarm Due:On and Off.
The selection will flash. See Figure 117.
5. Press
ENTER to confirm selection.
•
If you selected On, continue to step 6.
• If you select Off, pressing ENTER returns to the setup
mode.
Press
∆/MI to continue to the next setup program, or press
∇/MR to go to the previous program. Press EXIT to return to
mV measurement mode.
6. If you selected On in step 5, the display reads Key in Cal
Due Period (1-365 days) =. Key number of days using the
numeric keypad. If you make a mistake, use the
∆/MI or
∇/MR keys to highlight mistake, then re-enter.
See Figure 118.
7. Press
ENTER to confirm number of days to next calibration,
screen flashes Sys Updating, Please wait... and returns to
the setup mode.
Press
∆/MI to continue to the next setup program, or press
∇/MR to go to the previous program. Press EXIT to return to
mV measurement mode.
Keyin Cal Due Period
(1-365 days) = 1
Cal Alarm Due On Off
Select & Press ENTER
mVCalDue Alarm :P2.1
Press ENT. to Select
Figure 116
Figure 117
Figure 118