Instruction Manual

pH 1100/ 2100

39

10.3.3

mV calibration due alarm setup program P2.1


Use this program to set a reminder when the next calibration is
due. Choose from 1 to 365 days from last calibration.

1. Enter mV setup mode as described in steps 1-4 in Section

10.3.1 Entering mV setup mode (page 37).


2. Use

/MI key to scroll to program mVCalDue Alarm: P2.1.

See Figure 116.


3. Press

ENTER to enter program mVCalDue Alarm: P2.1.


4. Press

>/CAL to toggle between Cal Alarm Due:On and Off.

The selection will flash. See Figure 117.


5. Press

ENTER to confirm selection.


If you selected On, continue to step 6.

• If you select Off, pressing ENTER returns to the setup

mode.


Press

/MI to continue to the next setup program, or press

/MR to go to the previous program. Press EXIT to return to
mV measurement mode.


6. If you selected On in step 5, the display reads Key in Cal

Due Period (1-365 days) =. Key number of days using the
numeric keypad. If you make a mistake, use the

/MI or

/MR keys to highlight mistake, then re-enter.
See Figure 118.


7. Press

ENTER to confirm number of days to next calibration,

screen flashes Sys Updating, Please wait... and returns to
the setup mode.
Press

/MI to continue to the next setup program, or press

/MR to go to the previous program. Press EXIT to return to
mV measurement mode.

Keyin Cal Due Period
(1-365 days) = 1

Cal Alarm Due On Off
Select & Press ENTER

mVCalDue Alarm :P2.1
Press ENT. to Select

Figure 116

Figure 117

Figure 118