2 con 110 factory default setting – Thermo Fisher Scientific CyberScan CON 11/CON 110 User Manual

Page 74

background image

Instruction Manual

CON 11/110

70

14.2 CON 110 Factory Default Setting

PROGRAM FUNCTION

PARAMETER
OPTING KEYS

PARAMETER
OPTIONS

DEFAULT
SETTING

P 1.0

Unit Configuration

Parameter Title
Only

Parameter Title
Only

P 1.1

Selection of READY Indicator and Auto
HOLD Function

MI/▲; MR/▼

ON ; OFF,
HOLD

ON

P 1.2

Selection of °C and °F indicator

MI/▲; MR/▼

°C, °F

°C

P 1.3

Selection of Automatic or Manual
Temperature Compensation

MI/▲; MR/▼ YES,

nO YES

P 1.4

Selection of TDS Conversion Factor

MI/▲; MR/▼

0.40 to 1.00

0.50

P 2.0

Viewing previous calibration data

Parameter Title
Only

Parameter Title
Only

P 2.1

First range calibration point

Viewing Only

Viewing Only

_ _ _

P 2.2

Second range calibration point

Viewing Only

Viewing Only

_ _ _

P 2.3

Third range calibration point

Viewing Only

Viewing Only

_ _ _

P 2.4

Fourth range calibration point

Viewing Only

Viewing Only

_ _ _

P 2.5

Fifth range calibration point

Viewing Only

Viewing Only

_ _ _

P3.0

Viewing electrode data

Parameter Title
Only

Parameter Title
Only

P3.1

Effective cell constant for first range

Viewing Only

Viewing Only

K = “1.00”

P3.2

Effective cell constant for second range

Viewing Only

Viewing Only

K = “1.00”

P3.3

Effective cell constant for third range

Viewing Only

Viewing Only

K = “1.00”

P3.4

Effective cell constant for fourth range

Viewing Only

Viewing Only

K = “1.00”

P3.5

Effective cell constant for fifth range

Viewing Only

Viewing Only

K = “1.00”

P4.0:

Auto Power OFF

MI/▲; MR/▼ YES,

nO YES

P5.0

Reset to factory defaults

Parameter Title
Only

Parameter Title
Only

P5.1

Resetting of all calibration and clear all
calibration data

MI/▲; MR/▼ nO,

YES nO

P5.2

Resetting of all User settings to factory
defaults condition

MI/▲; MR/▼ nO,

YES nO

P6.0 Memory

clear

MI/▲; MR/▼ nO,

YES nO

P7.0 Temperature

Parameter Title
Only

Parameter Title
Only

P7.1

Setting of temperature coefficient

MI/▲; MR/▼

0.0 to 10.0

2.1

P7.2

Setting of normalisation temperature

MI/▲; MR/▼ 20.0,

25.0 25.0

P8.0

Selection of cell constant

MI/▲; MR/▼

1.0, 10.0, 0.1

K = “1.0”

P9.0 Automatic

Single Point Calibration

Parameter Title
Only

Parameter Title
Only

P9.1

Selection of Automatic or Manual
Calibration

MI/▲; MR/▼ YES,

nO YES

P9.2

Selection of Single or Multi Point
Calibration

MI/▲; MR/▼ YES,

nO YES

NOTE: The HOLD/ENTER key is used to confirm every parameter opted. The CAL/MEAS on the other hand will
escape the set up operation to return to the measurement mode.