Instruction Manual

CON 11/110

45

8.6

P5.0: Reset to factory defaults

Program 5.0 reset functions lets you have an option to only reset all the meter’s
calibrations and its data that have been stored or reset all calibrations, its stored
data, memory, and any other setup functions that you might have changed back to
factory default settings. See Factory Default Settings at page 69 and 70.

8.6.1 P5.1: Calibration Reset

This reset function clears all the calibrations and its data that
have been set by you. All other setup parameters that you
have changed are not reset.
From measurement mode

1. Press SETUP key to enter Set Up menu.
2. Press the MI/▲ or MR/▼ keys to until you view

parameter P5.0.

3. Press the HOLD/ENTER key to enter parameter

P5.1 with the upper display showing “CAL”.

4. Press the HOLD/ENTER key again to enter the

calibration reset selection mode.

5. Press the MI/▲ or MR/▼ keys to toggle between NO

and YES.

NO retains current settings

YES clears all calibrations and its data.

6. Press the HOLD/ENTER key to confirm selection

and to return to measurement mode. Otherwise
press CAL/MEAS key to return to measurement
mode without resetting the calibration data.


SETUP

cal

p 5.1

SETUP

cal

SETUP

SETUP

yes

MI

MR

HOLD

ENTER

HOLD

ENTER

p 5.0

cal

Figure 38: P5.1-Clear all

calibration and its data