8.8.2 P7.2: Setting the normalisation temperature

Your meter will normalise its conductivity measurements to a
standard temperature that you can select. You can adjust
the normalisation temperature either at 20.0°C or 25.0°C.
Meter default is 25.0°C.
From measurement mode

1. Press SETUP key to enter Set Up menu.
2. Press the MI/▲ or MR/▼ keys until you view

parameter P7.0.

3. Press the HOLD/ENTER key three times to select

display “tnr” at parameter P7.2.

4. Press the HOLD/ENTER key again. The upper

display shows the normalisation temperature of 25.0
°C.

5. Press the MI/▲ or MR/▼ keys to select either

20.0°C or 25.0°C as the normalisation temperature.

6. Press the HOLD/ENTER key to confirm selection

and to return to the subgroup menu.

7. Press CAL/MEAS key to return to measurement

mode.

SETUP

°C °F

p 7.2

p 7.2

p 7.2

25.0

SETUP

°C

20.0

SETUP

°C

MI

MR

HOLD

ENTER

Figure 42: P7.2-Setting

Normalisation Temperature