P 9.0, P 9.1 – Thermo Fisher Scientific CyberScan CON 11/CON 110 User Manual
Page 55

Instruction Manual
CON 11/110
51
8.10 P9.0: Automatic and Single point Calibration
You may select an automatic or manual calibration depending on the calibration
standards that you have. Program 9.0 lets you have the option to do the above as
well as the option of having a single or a multi point calibration.
8.10.1 P9.1: Selection of Automatic or Manual Calibration
The automatic calibration feature lets you calibrate the meter automatically by using
the appropriate known calibration standard solutions whereas in the manual
calibration feature non-standard calibration values can be used for calibration.
NOTE: Automatic calibration is applicable in conductivity
measurement mode only whereas manual calibration is
applicable in both conductivity and TDS mode.
Only from the conductivity measurement mode,
1. Press SETUP key to enter Set Up menu.
2. Press the MI/▲ or MR/▼ keys until you view
parameter P9.0.
3. Press the HOLD/ENTER key to enter parameter
P9.1. The upper display will show “ACA”.
4. Press the HOLD/ENTER key to enter the automatic
or manual calibration selection mode.
5. Press the MI/▲ or MR/▼ keys to select “YES” or
“NO”.
•
YES activates the automatic calibration.
•
NO activates the manual calibration.
6. Press the HOLD/ENTER key to confirm selection.
7. Press the CAL/MEAS key twice to return to
measurement mode.
NOTE: Parameter P9.1 will only be in the Set Up menu if
you enter from the conductivity measurement mode. If
you enter from the TDS measurement mode, Parameter
P9.1 will not be displayed.
Asc
SETUP
P 9.0
ACA
SETUP
P 9.1
ACA
SETUP
YES
ACA
SETUP
MI
MR
HOLD
ENTER
HOLD
ENTER
Figure 44: P9.1-Setting
Automatic or Manual
Calibration