8.10 P9.0: Automatic and Single point Calibration

You may select an automatic or manual calibration depending on the calibration
standards that you have. Program 9.0 lets you have the option to do the above as
well as the option of having a single or a multi point calibration.

8.10.1 P9.1: Selection of Automatic or Manual Calibration

The automatic calibration feature lets you calibrate the meter automatically by using
the appropriate known calibration standard solutions whereas in the manual
calibration feature non-standard calibration values can be used for calibration.
NOTE: Automatic calibration is applicable in conductivity
measurement mode only whereas manual calibration is
applicable in both conductivity and TDS mode.
Only from the conductivity measurement mode,

1. Press SETUP key to enter Set Up menu.
2. Press the MI/▲ or MR/▼ keys until you view

parameter P9.0.

3. Press the HOLD/ENTER key to enter parameter

P9.1. The upper display will show “ACA”.

4. Press the HOLD/ENTER key to enter the automatic

or manual calibration selection mode.

5. Press the MI/▲ or MR/▼ keys to select “YES” or

“NO”.

YES activates the automatic calibration.

NO activates the manual calibration.

6. Press the HOLD/ENTER key to confirm selection.
7. Press the CAL/MEAS key twice to return to

measurement mode.

NOTE: Parameter P9.1 will only be in the Set Up menu if
you enter from the conductivity measurement mode. If
you enter from the TDS measurement mode, Parameter
P9.1 will not be displayed.

Asc

SETUP

P 9.0

ACA

SETUP

P 9.1

ACA

SETUP

YES

ACA

SETUP

MI

MR

HOLD

ENTER

HOLD

ENTER

Figure 44: P9.1-Setting

Automatic or Manual

Calibration