beautypg.com

P 7.0, P 7.1 – Thermo Fisher Scientific CyberScan CON 11/CON 110 User Manual

Page 52

background image

Instruction Manual

CON 11/110

48

8.8 P7.0:

Temperature

8.8.1 P7.1:

Setting the temperature coefficient

The temperature coefficient is the amount of change in conductivity per degree of
temperature; it is expressed in percent per °C.. Entering the exact temperature
coefficient of your solution lets you accurately compensate
temperature for almost any solution. You can adjust 0.0 to
10.0 % per °C. Meter default is 2.1% per °C. By setting the
temperature coefficient to 0.0%, the meter will not apply any
compensation but will display actual temperature.
From measurement mode

1. Press SETUP key to enter Set Up menu.
2. Press the MI/▲ or MR/▼ keys until you view

parameter P7.0.

3. Press the HOLD/ENTER key to select parameter

7.1. The display shows “tCO” on the upper display.

4. Press the HOLD/ENTER key again. The upper

display shows the default temperature coefficient
“2.1”.

5. Press the MI/▲ or MR/▼ keys to select the

temperature coefficient of your solution.

6. Press the HOLD/ENTER key to confirm.
7. Press the CAL/MEAS key twice to return to the

measurement mode.

Notes
If you do not know the temperature coefficient of your
solution you can determine the correct value using the
formula in page 74 “Calculating Temperature Coefficients”.

SETUP

p 7.0

°C °F

SETUP

p 7.1

p 7.1

p 7.1

°C °F

2.1

SETUP

5.2

SETUP

HOLD

ENTER

HOLD

ENTER

MI

MR

Figure 41: P7.1 –

Setting Temperature

Coefficient