Instruction Manual

CON 11/110

50

8.9

P8.0: Selection of cell constant

Your meter lets you select a cell constant of K = 1.0, 10, or 0.1.
Use a cell of K = 1.0 for midrange measurements
Use a cell of K = 10 for high range measurements (above 20 mS or 10 ppt).
Use a cell of K = 0.1 for low range measurements (below 20
µS or 10 ppm).
The cell included with your meter has a cell constant of K =
1.0.
From measurement mode

1. Press SETUP key to enter Set Up menu.
2. Press the MI/▲ or MR/▼ keys until you view

parameter P8.0.

3. Press the HOLD/ENTER key to enter parameter

P8.0.

4. Press the MI/▲ or MR/▼ keys to select the cell

constant between K = 1.0, 0.1, or 10.

5. Press the HOLD/ENTER key to confirm selection

and to return to the subgroup menu.

6. Press the CAL/MEAS key to return to measurement

mode.



cel

1.0

SETUP

SETUP

p 8.0

p 8.0

p 8.0

10.0

SETUP

MI

MR

HOLD

ENTER

K =

K =

Figure 43: P8.0-

Selecting Cell Constant