P 4.0 – Thermo Fisher Scientific CyberScan CON 11/CON 110 User Manual
Page 48

Instruction Manual
CON 11/110
44
8.5
P4.0: Auto Off
In Program P4.0 you can activate the Auto Off option to automatically switch off the
instrument 20 minutes after the last key selection. This feature is useful for
conserving battery power. Under default conditions, the
instrument has this option activated.
1. Press the SETUP key to enter Set Up mode.
2. Press the MI/▲ or MR/▼ keys until you view
parameter P4.0.
3. Press the HOLD/ENTER key. The upper display
shows “AtO” and the lower display shows “YES” or
“NO”.
4. Press
MI/▲ or MR/▼ keys to select options
•
YES – activates the Auto Off function
•
NO - deactivates the Auto Off function
5. Press the HOLD/ENTER key to confirm and return
to subgroup menu.
6. Press the CAL/MEAS key to return to measurement
mode.
NOTE:
For datalogging using interval measurements, it is often desirable to use
an AC adapter and disable the auto-off feature.
ENTER
HOLD
SETUP
P 4.0
MR
MI
SETUP
SETUP
yes
O
O
O
Figure 37: P4.0-
Selection of Auto
Power Off function
- PCTestr 35 (2 pages)
- pHScan BNC (3 pages)
- pHScan 3/3+ (5 pages)
- pHTestr 1 (3 pages)
- pHTestr 10/20/30/10 BNC/Spear (2 pages)
- ORPTestr 10/10 BNC (2 pages)
- EC/TDS/SaltTestr 11 (4 pages)
- EC/TDS/SaltTestr (2 pages)
- ECScan High/Low & TDScan High/Low (9 pages)
- SaltTestr (2 pages)
- EcoTestr pH 2 (2 pages)
- EcoTestr pH 1 (2 pages)
- EcoTestr EC High (2 pages)
- EcoTestr EC Low (2 pages)
- EcoTestr TDS High (2 pages)
- EcoTestr TDS Low (2 pages)
- EcoTestr Salt (2 pages)
- Eutech pH 5/6 Plus & Ion 6 Plus (New version R1.1, SN >797406) (23 pages)
- Eutech pH 5/6 Plus & Ion 6 Plus (Old version EP6, SN <797406, discontinued) (23 pages)
- Eutech COND/TDS/Salt 6 Plus (40 pages)
- Eutech DO 6 Plus (48 pages)
- EcoScan pH/Ion 5 & 6 (27 pages)
- EcoScan CON 6 & TDS 6 (56 pages)
- EcoScan CON 5 & TDS 5 (18 pages)
- EcoScan Salt 6 (40 pages)
- EcoScan DO 6 (80 pages)
- CyberScan pH 10/pH 100 (67 pages)
- CyberScan pH 11/pH 110 (76 pages)
- CyberScan CON 10/CON 100/CON 200 (62 pages)
- CyberScan DO 110 (60 pages)
- CyberScan PCD 650 (127 pages)
- CyberScan CON 400/410 (For units manufactured before March 2010, discontinued) (60 pages)
- CyberScan CON 400 (For units manufactured from March 2010 onwards) (60 pages)
- CyberScan pH 300/310 (52 pages)
- CyberScan DO 300 (60 pages)
- CyberScan PC 300 (72 pages)
- CyberScan PD 300 (76 pages)
- CyberScan PC 10 (31 pages)
- C401 Colorimeter (64 pages)
- TN100 Turbidimeter (31 pages)
- RS232C Interface Adapter (9 pages)
- Thermo Scientific Temp 360 (44 pages)
- Thermo Scientific Temp 340 (40 pages)
- Thermo Scientific Temp 300 (32 pages)