8.4

P3.0 Viewing Electrode Data

Program 3.0 has five “view only” options that let you check the electrode’s
parameters for diagnostic purposes. These options show you the effective cell
constant for each range. The cell constant is adjusted
according to your calibration.
From measurement mode

1. Press the SETUP key to enter Set Up menu.

2. Press the MI/▲ or MR/▼ keys until you view

parameter P3.0.

3. Press

the

HOLD/ENTER

key repeatedly to view the

effective cell constant for each range.

4. When you have scrolled through all the electrode

data, you will automatically return to the subgroup
menu.

5. Press the CAL/MEAS key to return to measurement

mode.

NOTES: Cell constants will degrade with time and usage.
You can use this feature to alert you to the need for a new
electrode prior to total failure.

SETUP

1.00

p3.1

SETUP

p 3.0

ele

SETUP

p 3.2

1.00

ENTER

ENTER

K=

K=

Figure 36: P3.0-Viewing

electrode data for each

measurement range