P 1.4 – Thermo Fisher Scientific CyberScan CON 11/CON 110 User Manual

Page 45

background image

Instruction Manual

CON 11/110

41

8.2.4 P1.4: Selection of TDS Conversion Factor

You can set the TDS conversion factor between 0.40 and
1.00; meter default is 0.50.
NOTE: Your meter has to be in the TDS measurement mode
before you can make the selection of the TDS Conversion
Factor in the SETUP menu.
Only from TDS measurement mode

1. Press SETUP key to enter Set Up menu.
2. Press the HOLD/ENTER keys until you view

parameter P1.4.

3. Press the HOLD/ENTER key again. The upper

display shows a value and the lower display shows
“tdS”.

4. Press the MI/▲ or MR/▼ keys to select your

calculated TDS conversion factor.

5. Press the HOLD/ENTER key to confirm selection

and to return to the subgroup menu.

6. Press the CAL/MEAS key to return to measurement

mode.

ppt
ppm

SETUP

p 1.4

ppt
ppm

SETUP

0.50

ppt
ppm

SETUP

0.75

MI

MR

HOLD

ENTER

Figure 34: P1.4 -Select

TDS Conversion

Factor