Manual Temperature Compensation

To select the Manual Temperature Compensation (MTC) option, set up the Manual
Temperature Compensation mode in the SETUP menu Program P 1.3 by setting to
‘NO”.

1. Press SETUP key to enter Set Up mode. The display shows “COF” at “P

1.0”.

2. Press the HOLD/ENTER key to scroll through subgroups until you view

“ATC” at “P1.3” with small annunciators of “°C” and “ATC” or “°F” and “ATC”.

3. Press the HOLD/ENTER key again. The upper display shows “ATC” and the

lower display shows “YES”.

4. Press

the

MI/▲ or MR/▼ key to select the lower display to show “NO”.

5. Press the HOLD/ENTER key to confirm selection and to return to the

subgroup menu.

6. Press the CAL/MEAS key to return to measurement mode.

CAL

MEAS

SETUP

YES

°C

ATC

SETUP

°C

HOLD

ENTER

MI

MR

HOLD

ENTER

SETUP

cof

p 1.0

No ‘ATC’

SETUP

READY

ON

p 1.1

SETUP

C

p 1.2

°C

SETUP

A C

p 1.3

°C

HOLD

ENTER

HOLD

ENTER

HOLD

ENTER

SETUP

cof

p 1.0

SETUP

Figure 16: Manual Temperature

Compensation Set Up sequence