Instruction Manual

CON 11/110

36

8.1

Advanced SETUP Mode Overview

1. Press the SETUP key to enter Set Up menu.
2. Press

the

MI/▲ or MR/▼ keys to scroll through sub groups.

3. Press HOLD/ENTER key to enter a particular parameter.

SETUP

P 1.0

COF

P1.0: Unit Configuration
P1.1 Selection of READY and *Auto HOLD function (*CON
110

Only).

P1.2 Selection of °C or *°F (*CON 110 Only)
P1.3 Selection of Automatic or Manual Temperature
Compensation
P1.4 Selection of TDS Conversion Factor (This mode only

appears when SETUP Menu is chosen while in the

TDS measurement mode)

SETUP

CAL

P 2.0

P2.0: Viewing previous calibration data
P2.1 First range calibration point
P2.2 Second range calibration point
P2.3 Third range calibration point
P2.4 Fourth range calibration point
P2.5 Fifth range calibration point

SETUP

ELE

P 3.0

P3.0: Viewing electrode data
P3.1 Effective cell constant for first range
P3.2 Effective cell constant for second range
P3.3 Effective cell constant for third range
P3.4 Effective cell constant for fourth range
P3.5 Effective cell constant for fifth range

SETUP

P 4.0

A o

P4.0: Auto Off
P4.0 Selection

of

automatic power off function

SETUP

P 5.0

P5.0: Reset to factory defaults
P5.1 Resetting of all calibration and clear all
calibration

data

P5.2 Resetting of all User settings to factory
defaults

condition

SETUP

P 6.0

MEM

P6.0: Memory clear
P6.0 Clearance of all stored readings

SETUP

P 7.0

°C °F

P7.0: Temperature
P7.1 Setting of temperature coefficient
P7.2 Setting of normalisation temperature