Instruction Manual

CON 11/110

20

5 MEASUREMENT

This meter is capable of taking measurements with automatic (ATC) or manual
(MTC) temperature compensation. Factory default is ATC on.

5.1 Automatic

Temperature

Compensation

For automatic temperature compensation (ATC), simply attach
the conductivity/TDS electrode to the meter. Ensure that the
ATC mode has been selected in the SETUP menu. The ATC
indicator will light on the LCD.
NOTE: If the conductivity/TDS electrode is not properly
attached to the meter or it has been damaged, the ATC
indicator will blink and the temperature display will show
“Ur”.
If the ATC indicator does not light and the temperature
display shows a reading, ATC mode may have been replaced with the manual
temperature compensation mode in the meter’s SETUP menu. See page 21 Section
5.2 on Manual Temperature Compensation.
Automatic Temperature Compensation mode can be restored by setting the SETUP
Menu Program P1.3 to YES.

MEAS

READY

1413

ATC

°C

22.3

µS

Figure 15: Automatic

Temperature Compensation