8.6.2 P5.2: User Reset

This reset function clears all the calibrations, memory and
all other setup parameters that you have changed back to
factory defaults condition.
From measurement mode

1. Repeat step 1 to 4 to enter the calibration reset

selection mode.

2. Press

the

MI/▲ or MR/▼ keys to select NO.

3. Press the HOLD/ENTER key to enter parameter

P5.2. The upper display will show “USr”.

4. Press the HOLD/ENTER key again to enter user

reset selection mode.

5. Press the MI/▲ or MR/▼ keys to toggle between

NO and YES.

NO retains current settings

YES resets to factory default settings.

6. Press the HOLD/ENTER key after confirming reset

selection and to return to the measurement mode.
Otherwise press CAL/MEAS key to return to
measurement mode without resetting to factory
default.

NOTE: See page 69 and 70 for Factory Default
Settings.





SETUP

cal

yes

SETUP

SETUP

SETUP

P 5.2

ENTER

HOLD

ENTER

HOLD

MI

MR

Figure 39: P5.2-Clear

all setups to factory

default