11 ERROR

MESSAGES

The following table provides a guideline to enable diagnosis of possible problems
indicated by the messages generated by the meter. The table also provides possible
solutions to the problems encountered.

ERROR MESSAGE

INDICATES

POSSIBLE CAUSE

CORRECTIVE
ACTION

ERR. Annunciator

Wrong keypad input

Wrong input in
selected mode.

Release key. Select
valid operations
depending on mode.

Secondary Display
Toggles Continuously
between Calibration
Standard Values

Calibration error

Incorrect calibration
solution used or
contamination of
calibration solution
occurring in automatic
calibration mode

Check if the right
calibration solution
was selected or use
fresh calibration
solution.

Battery icon lights up

Low Battery

Battery power is low

Replace batteries
with fresh ones as
soon as possible.