beautypg.com

P 1.4, P 1.0 – Thermo Fisher Scientific CyberScan CON 11/CON 110 User Manual

Page 22

background image

Instruction Manual

CON 11/110

18

4.7 TDS

Calibration

4.7.1 Calibrating for TDS Using Conductivity Standards & adjusting TDS factor

It is important to note that the increase in concentration of salts dissolved in solution
increases the conductivity of that solution. This relationship varies from salt to salt
and is roughly linear over a given range for a given salt. The TDS conversion factor is
the number used by the meter to convert from conductivity to TDS.
Instead of calibrating for TDS directly using TDS calibration standard solutions, you
can have TDS calibration by using the conductivity calibration method and enter the
appropriate TDS conversion factor into the meter.
For more information regarding TDS Conversion Factor determination, please refer to
page 73.

4.7.2 Setting the TDS Conversion Factor

The factory default setting for TDS conversion factor is 0.5.
If your solution has a different TDS factor, you can improve
calibration accuracy by setting the TDS factor prior to
calibration.

1. Press MODE key to select TDS mode.
2. Press Setup key to enter SETUP Menu.
3. Press

the

HOLD/ENTER

key a few times to scroll

through the sub programs until you view Program
P1.4

4. Press the HOLD/ENTER key again. The upper

display shows a value and the lower display shows
“tdS”.

5. Press the MI/▲ or MR/▼ keys to select your

calculated TDS conversion factor.

6. Press the HOLD/ENTER key to confirm selection

and to return to the subgroup menu.

7. Press the CAL/MEAS key to return to

measurement mode.

NOTE: The set up program P1.4 for setting the TDS
conversion factor is only activated when set up menu is
entered while the meter is in the TDS measurement mode.




ppt
ppm

SETUP

p 1.4

ppt
ppm

SETUP

0.50

ppt
ppm

SETUP

0.75

SETUP

cof

p 1.0

SETUP

MI

MR

HOLD

ENTER

HOLD

ENTER

Only from TDS Measurement Mode

Figure 14: Setting the TDS

Conversion Factor