1/3

1.3 Electrical connection of separate primary head (F Versions)

1.3.1 General information on signal cable A and field current cable C

Use of the Krohne signal cable A with foil screen and magnetic shield will ensure proper operation
of the equipment.

• Signal cable to be solidly laid.
• Connect shields via stranded drain wires.
• Underwater and underground installation possible.
• Insulating material is flame-retardant to IEC IEC 332.1 / VDE 0742.
• Signal cables are low in halogen, unplasticized, and stay flexible at low temperatures.

Signal cable A (Type DS), with double shielding

1

Stranded drain wire, 1st shield, 1.5 mm

2

or AWG14

2

Insulation

3

Conductor 0.5 mm

2

or AWG 20 (3.1 red / 3.2 white)

4

Special foil, 1st shield

5

Inner sheath

6

Mu-metal foil, 2nd shield

7

Stranded drain wire, 2nd shield, 0.5 mm

2

or AWG 20

8

Outher sheath

Field current cable C with single shielding
Cross-section is dependent on required length of cable, see Table in Sect. 1.3.4.

1.3.2 Stripping (preparation) of signal cable A

Please note the different lengths given in the table for signal converter and primary head.

Length Converter

Primary
head

mm (inch)

mm (inch)

a

70

(2.80)

90

(3.60)

b

08

(0.30)

08

(0.30)

c

25

(1.00)

25

(1.00)

d

08

(0.30)

08

(0.30)

e

50

(2.00)

70

(2.80)

Customer-supplied materials
W
Insulation tubing (PVC), 2.0 - 2.5 mm dia. (1”)
X Heat-shrinkable tubing or cable sleeve
Y Wire end sleeve to DIN 41 228: E 1.5-8
Z

Wire end sleeve to DIN 41 228: E 0.5-8

Please note:
For primary heads, stranded drain wire 1
must have the same length as stranded drain wire 7.

See Section 1.3.4 for max. permissible cable lengths

Signal cable A

bending radius

50 mm (

2”)