Versions

K = contact

F = separate, field housing

IFC 090 K/B and F/B (standard)

Basic version, without local display and control elements

IFC 090 K/D and F/D (option)

Display version, with local display and control elements

IFC 090 K/D-EEx

Hazardous-duty version with outputs in Increased Safety

Interface (option)

HART

Add-on equipment (option)

CONFIG software and adapter for operator control via
MS-DOS PC, connection to internal IMoCom interface
(equipment bus)

Current output
Function

– all operating data settable
– galvanically isolated from all output and input circuits
– for active or passive mode (Ex version only active)

Current:

fixed ranges

0 – 20 mA and 4 – 20 mA

variable ranges

for Q = 0%

I

0%

= 0 – 16 mA

adjustable in 1mA

for Q = 100%

I

100%

= 4 – 20 mA

increments

for Q > 100%

I

max

> 20 up to 22 mA maximum

Active mode

max. 500

load

Passive mode

external voltage:

15 ... 20 V DC

20 ... 32 V DC

load: min. ... max.

0 ... 500

250 ... 750

Error identification

0 / 22 mA and variable

Forward/reverse flow measurment

direction identified via status output

Pulse output
Function

– all operating data settable
– galvanically isolated from current output and all input circuits
– digital pulse division, interpulse period non-uniform,

therefore if frequency and cycle meters connected
allow for minimum counting interval:

gate time, totalizer

Activ mode

connection: electronic totalizers
voltage: approx. 15 V DC, from current output
load: I

max

< 23 mA, operation without current output

load:

I

max

<

2

3 mA, operation with current output

Passive mode

connection: electronic or electromechanical totalizers
voltage: external, U

ext

32 V DC/

24 V AC

load: I

max

150 mA

Pulse width

automatic: pulse duty cycle 1:1, max. 1000 pulses/s = 1 kHz

variable: 10 ms – 1 s P

100%

[pulses/s] = f

max

[Hz] =

Forward/reverse flow measurement

flow direction identified via status output

Status output (passive)
Function

settable as measuring range identification for BA mode,
indicator for flow direction, errors or trip point

Connection

voltage: external, U

ext

32 V DC/

24 V AC

load current: I

max

150 mA

10/3

10.4

IFC 090 Signal converter

}

1000

P

100%

[Hz]

1

2x pulse width