F/R flow measurement

overranging
of outputs

}

dynamic response
of outputs, see Fct. 1.02
TIMECONST.:
I = I only
P = ALL

1.06

CONTROL B1

Control input B1 and B2 (see Fct. 3.07 HARDWARE)

1.07

CONTROL B2

OFF

EXT.RANGE (external range change)

Setting range: 5 - 80 PERCENT (= ratio of lower to upper range from

1:20 to 1:1.25. Value must be greater than that of Fct. 1.03 L.F. CUTOFF).

Press

key to transfer to number setting.

OUTP.HOLD (hold value of outputs)
OUTP.ZERO (set outputs to “min.values“)
TOTAL. RESET (reset totalizers)
ERROR. RESET (delete error messages)

Press key

to return to

Fct. 1.06 or 1.07 CONTROL B1 or B2.

Fct.

Text

Description and settings

1.06

PULS B1

Pulse output B1 (see Fct. 3.07 HARDWARE)

FUNCT. P

Select function for pulse output P
OFF (switched off)
1 DIR. (1 flow direction)
2 DIR. (forward/reverse flow, F/R flow measurement)

Press key

to transfer to subfunction

“SELECT P”.

SELECT P

Select pulse type
PULSE/VOL. (pulses per unit volume, flowrate)
PULSE/TIME (pulses per unit time for 100% flowrate)

Press key

to transfer to subfunction

“PULSWIDTH”.

PULSWIDTH

Select pulse width
0.01 - 1.00 Sec (only for F

max

< 50 pulses/s)

AUTO (automatic = 50% of the period duration

of the 100% output frequency)

SYM. (symmetrical = pulse duty ratio 1:1 over total range)

Press key

to transfer to subfunction

“VALUE P”.

VALUE P

Set pulse value per unit volume (appears only when
“PULSE/VOL.” set under “SELECT P” above)
xxxx PulS/m3 xxxx PulS/Liter xxxx PulS/US.Gal
xxxx PulS/ user-defined unit, factory-set is “Liter” or “US M.Gal” (see Fct. 3.05)
Setting range “xxxx” is dependent on the pulse width and the
full-scale range: P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Press key

to return to

Fct. 1.06 “PULS. B1”.

VALUE P

Set pulse value per unit time (appears only when “PULSE/TIME”
set under “SELECT P” above)
xxxx PulSe/Sec (=Hz) xxxx PulSe/min xxxx PulSe/hr
xxxx PulSe/user-defined unit, factory-set is “hr” or “day” (see Fct. 3.05)
Setting range “xxxx” is dependent on the pulse width, see above

Press key

to return to

Fct. 1.06 “PULS. B1”.

1.06

STATUS B1

Status output B1 and B2 (see Fct. 3.07 HARDWARE)

1.07

STATUS B2

ALL ERROR

FATAL ERROR

OFF

ON

• SIGN. I
• SIGN. P

}

OVERFLOW I
OVERFLOW P

}

AUTO RANGE (automatic range change)

Setting range 5 - 80 PERCENT (= ratio of lower to upper range from
1:20 to 1:1.25. Value must be higher than that of Fct. 1.03 L.F. CUTOFF)

TRIP POINT:

XXX

- YYY

XXX > YYY N/O contact

0 - 150% 0 - 150% XXX < YYY N/C contact
Hysteresis

1%

(difference between XXX value and YYY value)

EMPTY PIPE (signals that pipe is “empty”, only if option installed)

Press key

to transfer to number setting.

Press key

to return to

Fct. 1.06 or 1.07 STATUS B1

or

B2.