If you need to return your flowmeter to Krohne,

please refer to last but one page of these Instructions !

Example

Full-scale range

Q

100%

= 200 m

3

/hr (Fct. 1.01)

Meter size

DN

= 80 mm = 3“ (Fct. 3.02)

Current at Q

0%

I

0%

= 4 mA

Q

100%

I

100%

= 20 mA

}

(Fct. 1.05)

Pulses at

Q

100%

P

100%

= 280 pulses/hr (Fct. 1.06)

Primary head constant

GK

= 3.571 (see nameplate)

Constant

(V

in m3)

(t

in hr)

K

= 7074 (see Table)

(DN in mm)

Calculation of “ X ” and position of “ Y ”

X =

Q

100%

x K

=

200 x 7074

= 61.905

GK x DN

2

3.571 x 80 x 80

Y = 80, position of switch Y, see front panel GS 8A
(comes closest to X value and is smaller than X).

Calculation of setpoint readings I and f

I = I

0%

+

Y

(I

100%

- I

0%

) = 4 mA +

40

(20mA - 4mA) = 14.3 mA

X 61.905

Deviations are permissible between 14.1 and 14.6 mA (equivalent to ± 1.5 %).

f =

Y

x P

100%

=

40

x pulses / hr = 129.2 pulses/hr

X 61.905

Deviations are permissible between 127.3 and 131.1 pulses/hr (equivalent to ± 1.5 %).