KROHNE IFC 090 Converter EN User Manual
Page 37
5/6
→
VALUE P = set pulse value per unit volume
(appears only when “PULSE/VOL.” set under “SELECT P”, press
→
key.
•
XXXX PulS/m3
•
XXXX PulS/Liter
•
XXXX PulS/US.Gal
•
XXXX PulS/ user unit, factory-set: “Liter” or “US MGal”, see Sect. 5.13.
Select using
↑
key.
Transfer to number setting with
→
key. 1st digit (cursor) flashes.
Set numerical value
•
XXXX
(setting range depends on pulse width and
full-scale range: P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
)
Change flashing digit (cursor) with
↑
key,
shift cursor 1 place to right or left with
→
key.
Press
↵
key to return to Fct. 1.06 PULS B1.
or
Please refer to Sect. 3.2 “factory settings”
→
VALUE P = set pulse value per unit time,
(appears only when ”PULSE/TIME” has been set under “SELECT P”),
press
→
key.
•
XXXX PulSe/Sec
•
XXXX PulSe/min
•
XXXX PulSe/hr
•
XXXX PulSe/ user unit, factory-set: “hr”, or “day”, see Sect. 5.13.
Select using
↑
key.
Transfer to number setting with
→
key, 1st digit (cursor) flashes.
Set numerical value
•
XXXX
(setting range depends on pulse width)
Change flashing digit (cursor) with
↑
key,
shift cursor 1 place to right or left with
→
key.
Press
↵
key to return to Fct. 1.06 PULS B1.
Refer to Sect. 2.6 for connection diagrams, and to Sect. 5.15 for characteristics.