KROHNE IFC 090 Converter EN User Manual
Page 33
5/2
Low-flow cutoff 5.3
Display 5.4
Fct. 1.04 DISPLAY
Press
→
key.
→
DISP. FLOW = select unit for display of flowrate,
press
→
key
•
NO DISP.
(no display)
•
m3/hr
(cubic metres per hour)
•
Liter/Sec
(litres per second)
•
US.Gal/min
(US gallons per minute)
•
user-defined unit, factory-set: “Liter/hr” (litres per hour) or “US MGal/day”, see Sect. 5.15
•
PERCENT
(percentage display)
•
BARGRAPH
(numerical value and bar graph display in %)
Select with
↑
key.
Press
↵
key to transfer to subfunction “DISP. TOTAL”.
→
DISP. TOTAL = select unit for totalizer display,
press
→
key
•
NO DISP.
(no display)
•
OFF
(internal totalizer switched off)
• + TOTAL. • – TOTAL. • +/– TOTAL. • SUM (
ΣΣ
) • ALL (sequential)
Select with
↑
key.
Transfer to totalizer unit setting using
↵
key.
•
m3
(cubic metres)
•
Liter
(litres)
•
US.Gal
(US gallons)
•
user-defined unit, factory-set: “Liter” or “US MGal”, see Sect. 5.12
Select with
↑
key.
Transfer to totalizer format setting using
→
key.
Continuation see next page
Fct. 1.03 L.F.CUTOFF
Press
→
key.
Choice
•
OFF
(fixed tripping point:
ON = 0.1 % / OFF = 0.2 %)
•
PERCENT
(variable tripping points: ON = 1 - 19 % / OFF = 2 - 20 %)
Select with
↑
key.
Transfer to number setting using
→
key (only if “PERCENT” selected).
1st number (cursor) flashes.
Setting the numerical value when “PERCENT” selected
•
01 to 19
(cutoff “on” value, left of hyphen)
•
02 to 20
(cutoff “off” value, right of hyphen)
Change flashing number (cursor) with the
↑
key.
Shift cursor 1 place to right using
→
key.
Press
↵
key to return to Fct. 1.03 L.F.CUTOFF.
Note: The cutoff “off” value must be greater than the cutoff “on” value.