Control input (passive)
Function

– settable for range change, totalizer reset, error reset,

set outputs to min. values or hold actual output values

– initiate function by “low” or “high” control signals

Control signals

U

max

: 24 V AC

32 V DC (any polarity)

low:

1,4 V

2 V

high:

3 V

4 V

Output/input combinations

I = current output

P = pulse output

S = status output

C = control input

the following combinations can be set:
1) I

P

S

2) I

P

C

3) I

C

S

4) I

S

C

5) I

S1

S2

6) I

C1

C2

Time constant

0.2 – 99.9 s, adjustable in increments of 0.1 second

Low-flow cutoff

Cutoff “on” value: 1 – 19% of Q

100%

, adjustable

Cutoff “off” value: 2 – 20% in 1% increments

Local display (D version)

3-field LCD

Display function

actual flowrate, forward, reverse and sum totalizers
(7-digit), or 25-character bar graph with percentage
indication and status messages

Units:

Actual flowrate

m

3

/h, liter/s., US gallons/min or user-defined unit,

e. g. liters/day

Totalizer

m

3

, liters, US gallons or user-defined unit,

e. g. hecto-liters or US gallons/day
(adjustable count duration up to overflow)

Language of plain texts

English, German, French, others on request

Display:

Top field

8-character, 7-segment numeral and sign display,
and symbols for key acknowledgement

Middle field

10-character, 14-segment text display

Bottom field

4 markers to identify display in measuring mode

Operator control

– 3 keys

→ ↵ ↑

or

– 3 magnetic sensors and the supplied

bar magnet without opening the housing

Power supply

1. Rated voltage
Tolerance band

2. Rated voltage
Tolerance band

Frequency

Power consumption
(incl. primary head)

Housing
Material

die-cast aluminium with polyurethane finish

Ambient temperature

– 25 to + 60°C (– 13 to + 140°F)

Protection category (IEC 529/EN 60 529)

IP 67, equivalent to NEMA 6

}

1. AC Version

2. AC Version

AC/DC Version

Standard

Option

Option

230 / 240 V

200 V

24 V AC

24 V DC

200 – 260 V

170 – 220 V

20 – 27 V AC

18 – 32 V DC

115 / 120 V

100 V

100 – 130 V

85 – 110 V

48 – 63 Hz

48 – 63 Hz

approx. 10 VA

approx. 10 VA

approx. 8 W

When connected to functional extra-low voltage, 24 V, safety separation (PELV)
is essential (to VDE 0100/VDE 0106 and IEC 364/IEC 536 or equivalent
national standard).