beautypg.com

KROHNE IFC 090 Converter EN User Manual

Page 31

background image

4/12

The cursor, flashing part of display, is shown below in bold type.

Change measuring range of current output and value for error messages (Fct. 1.05):

Change measuring range from 04-20 mA to 00-20 mA

Change value for error messages from 0 mA to 22 mA

Key

Display

Description

If ”YES“ set under Fct. 3.04 ENTRY CODE, key in the
9-keystroke CODE 1 now:

→ → → ↑ ↑ ↑ ↵ ↵ ↵

Fct. 1.00

OPERATION

Fct. 1.01

FULL SCALE

4x

Fct. 1.05

CURRENT I

FUNCT. I

→ ↵

RANGE I

If ”REV. RANGE“ appears here,
press keys

and

again.

04-20

mA

Old current range

2x

00-20

mA

New current range

I ERROR

0

mA

Old value for error messages

22

mA

New value for error messages

Fct. 1.05

CURRENT I

Fct. 1.00

OPERATION

STORE YES

- - - - - - -

- - - - - / - - -

Measuring range with new data for the current output

Examples of setting the signal converter

4.7