KROHNE IFC 090 Converter EN User Manual

Page 56

background image

7/5

Part 2

Signal converter IFC 090 D (D = display version) and
Signal converter IFC 090 B (B = basic version), without display
but with HHT or CONFIG operator program (see Sect. 6.2)

Group D

Display shows . . .

Caus

Remedial action

D 1

LINE INT.

Power failure.

Delete error message in

Note: no counting during

RESET/QUIT. menu. Reset

power failure.

totalizer(s) if need be.

D 2

CUR.OUTP. I

Current output overranged.

Check instrument parameters
and correct if necessary.
Error message deleted
automatically after cause
has been eliminated.

D 3

PULS.OUTP. P

Pulse output overranged.

Check instrument parameters,

Note: totalizer deviation

correct if necessary, and reset

possible

totalizer(s). Error message
deleted automatically
after cause has been
eliminated.

D 4

ADC

Analog/digital converter

Error message deleted

overranged.

automatically after cause
has been eliminated.

D 5

FATAL. ERROR

Fatal Error,

Replace signal converter

all outputs set to “min.”

(see Sect. 8.7) or contact

values.

Krohne Service, having
first noted down hardware
information and error status,
see Fct. 2.02.

D 6

TOTALIZER

Counts lost

Delete error message in

(overflow, data error)

RESET/QUIT. menu.

D 7

„STARTUP“

Hardware fault,

Replace signal converter

cyclic flashing

Watchdog activated.

(see Sect. 8.7) or contact
Krohne Service.

D 8

BUSY

Displays for flow,

Change setting in

totalizers and errors

Fct. 1.04.

disabled.

D 9

Unsteady display

Low electrical

Increase time constant in

conductivity, high solids

Fct. 1.02.

content, pulsating flow.

D 10

No display

Power OFF.

Switch on power

Check power (fuse(s)

Replace if defective

F1 (F1 + F2 with DC).

(see Sect. 8.1).