5/4

Current output I 5.6

Fct. 1.05 CURRENT I
Press

key.

FUNCT. I = Select function for current output,

press

key

OFF

(switched off, no function)

1 DIR.

(1 flow direction)

2 DIR.

(2 flow directions, F/R mode, forward/reverse)

Select using

key.

Transfer to subfunction “RANGE I” with

key.

Exceptions: When “OFF” selected, return to Fct. 1.05 CURRENT I.

When “2 DIR.”selected, transfer to subfunction “REV.RANGE”.

REV.RANGE = define full-scale range for reverse flow

(appears only when “2 DIR.” set under “FUNCT. I” above)
Press

key

100 PCT. (same full-scale value Q

100%

as forward flow, see Fct. 1.01)

PERCENT (settable range) Setting range 005 – 150% of Q

100%

(see Fct. 1.01)

Select with

key.

Press

key to transfer to number setting.

Press

key to transfer to subfunction “RANGE I”.

RANGE I = select measuring range,

press

key

0 - 20 mA

fixed ranges

4 - 20 mA

mA (user-defined value) I

0%

– I

100%

(value I

0%

< I

100%

!) 0-16 mA 4-20 mA

Press

key to transfer to number setting.

Select with key

.

Press key

to transfer to subfunction “I ERROR”.

I ERROR = set error value,

press

key

22 mA

(fixed value)

0.0 - I

0%

mA (variable value; only variable when I

0%

1 mA, see ”RANGE I“ above)

Select using key

. Press

key to transfer to number setting.

Press key

to return to Fct. 1.05 CURRENT I.

Please refer to Sect. 3.2 “Factory settings”.

Refer to Sect. 2.6 for connection diagrams, and to Sect. 5.15 for characteristics.

}