KROHNE IFC 090 Converter EN User Manual

Page 28

background image

4/9

Fct.

Text

Description and settings

3.03

ZERO SET

Zero calibration
Note:

Carry out only at “0” flow and with completely filled
measuring tube!

Precautionary query
CALIB. NO

Press

key to return to Fct. 3.03 “ZERO SET”.

CALIB. YES

Press

key to start calibration.

Duration approx. 25 seconds, current flowrate
displayed in the selected unit (see Fct. 1.04 “DISP. FLOW”).

A “WARNING” sign appears when flowrate “>

0

“;

acknowledge by pressing

key.

STORE NO (do

not store new zero value)

STORE YES (store new zero value)

Press

key to return to Fct. 3.03 “ZERO SET”.

3.04

ENTRY CODE

Entry code required to enter setting mode?
NO

(= entry with

only)

YES (= entry with

and Code 1:

→ → → ↵ ↵ ↵ ↑ ↑ ↑

)

Press

to return to Fct. 3.04 “ENTRY CODE”.

3.05

USER UNIT

Set any required unit for flowrate and counting

TEXT VOL.

Set text for required flowrate unit (max. 5 characters)
Factory-set: “Liter” or “MGal”.
Characters assignable to each place:
A-Z, a-z, 0-9, or “ – “ (= blank character).

Press

key to transfer to subfunction “FACT. VOL.”

FACT. VOL.

Set conversion factor (F

M

) for volume

Factory set “1.00000” for “Liter” or “2.64172E-4” for “US MGal”
(exponent notation, here: 1

x

10

3

or 2.64172

x

10

-4

).

Factor F

M

= volume per 1m

3

.

Setting range

1.00000 E-9 to 9.99999 E+9 (= 10

-9

to 10

+9

)

Press

key to transfer to subfunction “TEXT TIME”.

TEXT TIME

Set text for required time unit (max. 3 characters)
Factory-set: “hr” or “day”:
Characters assignable to each place:
A-Z, a-z, 0-9, or “ – “ (= blank character).

Press

key to transfer to subfunction “FACT. TIME”

FACT. TIME

Set conversion factor (F

T

) for time

Factory-set: “3.60000 E+3” for “hour” or “8.64000 E+4” for “day”
(exponent notation, here: 3.6

x

10

3

or 8.64

x

10

-4

).

Set factor F

T

in seconds.

Setting range
1.00000 E-9 to 9.99999 E+9 (= 10

-9

to 10

+9

)

Press

key to return to Fct. 3.05 “USER UNIT”.

3.06

APPLICAT.

Set overload point for A/D converter

FLOW

STEADY (150% of Q

100%

) PULSATING (1000% of Q

100%

)

Press key

to return to

Fct. 3.06 “APPLICAT.“,

with installed option

“empty pipe“

,

change to subfunction

“EMPTY PIPE“.

EMPTY PIPE

Switch on “empty pipe” identifier option?
(appears only when this option is installed)
YES NO

Select with key

.

Press

key to return to

Fct. 3.06 “APPLICAT.“.

3.07

HARDWARE

Assign outputs and inputs to terminals B1 and B2

TERM.B1

Terminal B1
PULSOUTP. STATUSOUTP. CONTROLINP.

Select with key

.

Press key

to transfer to subfunction

“TERM. B2”.

TERM.B2

Terminal B2
STATUSOUTP. CONTROLINP.

Select with key

.

Press key

to return to

Fct. 3.07 “HARDWARE”.