7.4 Faults and symptoms during start-up and process flow measurement
Before contacting Krohne Service, please read through the
following tables. THANK YOU !
•
Most faults and symptoms occuring with the flowmeters can be eliminated
with the aid of the following tables.
•
For greater clarity, faults and symptoms in the tables are divided into 2 parts and
various groups.
•
Part 1
Signal converter IFC 090 B (B = basic version), without display
and without HHT or CONFIG user program (see Sect. 6.2)
Groups:
LED
LED display (status messages)
I
Current output
P
Pulse output
LED / I / P LED display, current output and pulse output
•
Part 2
Signal converter IFC 090 D (D = display version) and
Signal converter IFC 090 B (B = basic version), without display
but with CONFIG user program (see Sect. 6.2)
Groups:
D
Display
I
Current output
P
Pulse output
S
Status indication output
C
Control input
D / I / P / S LED display, current output, pulse output,
status output and display
Part 1
Converter IFC 090 B (B = basic version), without display
and without HHT or CONFIG operator program
Group LED
Fault / Symptom
Cause
Remedial action
LED 1
LED flashes red/green
Overranging of A/D converter,
Reduce flowrate; if unsuccessful,
current or pulse output
test as described in Sect. 7.5.
Measuring tube drained,
Fill measuring tube.
A/D converter overranged.
LED 2
LED flashes red
Fatal Error, hardware and/or
Replace signal converter
software error
(see Sect. 8.7) or contact
Krohne Service.
LED 3
cyclic flashing of red LED,
Hardware fault, Watchdog
Replace signal converter
approx 1 second
activated.
(see Sect. 8.7) or contact
Krohne Service.
LED 4
LED shows red continuously
Hardware fault
Replace signal converter
(see Sect. 8.7) or contact
Krohne Service.
Group I
Fault / Symptom
Cause
Remedial action
I 1
Receiver instrument
Connection/polarity incorrect.
Connect properly as described
indicates “0”.
in Sect. 2.3 + 2.7
Receiver instrument
Check connecting cables and
defective.
receiver instrument, and
replace if necessary.
Short between current output
Check connections and
and pulse output
cables, see Sect. 2.3 + 2.7,
voltage between I+ and
I
⊥
approx. 15 V. Switch off
device, eliminate short-circuit,
switch device on.
Current output defective
Replace signal converter
(see Sect. 8.7) or contact
Krohne Service.
I 2
22 mA present at current
Current output I overranged
Check device parameters
output (fault current)
and change if necessary,
see Sect. 6.2, or contact
Krohne Service