Signal converter IFC 090 D (D = display version) and
(cont’d)
Signal converter IFC 090 B (B = basic version), without display
but with HHT or CONFIG operator program (see Sect. 6.2)
Group S
Fault / Symptom
Cause
Remedial action
S 1
No function
Incorrect connection/polarity
Connect properly,
of status display
see Sect. 2.5 + 2.7.
Status display or output
Set status output under Fct. 1.07
defective or external
to “F/R INDIC.” (flow direction)
voltage source not
and check (see Sect. 7.2)
supplying voltage.
with new status display:
Test ok, check previous
status display and external
voltage source, and replace
if necessary.
Test faulty, status output
defective, replace signal
converter (see Sect. 8.7) or
contact Krohne Service.
Output terminal B1 or B2
Change setting under
not set to “status output”
Fct. 3.07
Group D/I/P/S
Fault / Symptom
Caus
Remedial action
D / I / P / S 1
Unsteady display and
Electrical conductivity
Increase time constant under
outputs
of product too low,
Fct. 1.02.
time constant too low.
D / I / P / S 2
No display and no
Power OFF
Switch power on.
function of outputs
Check power fuse(s)
Replace if defective,
F1 (F1 + F2 for DC).
see Sect. 8.1.
Group C
Fault / Symptom
Cause
Remedial action
C 1
No function
Connection is incorrect
Connect properly,
see Sect. 2.6 + 2.7
Control input C
Check connection, cables
or external voltage
and external voltage source,
source defective
see Sect. 2.6 + 2.7
Output terminal
Change setting under
B1 or B2 not set to
Fct. 3.07.
“control input”