Part 2

Signal converter IFC 090 D (D = display version) and

(cont’d)

Signal converter IFC 090 B (B = basic version), without display
but with HHT or CONFIG operator program (see Sect. 6.2)

Group S

Fault / Symptom

Cause

Remedial action

S 1

No function

Incorrect connection/polarity

Connect properly,

of status display

see Sect. 2.5 + 2.7.

Status display or output

Set status output under Fct. 1.07

defective or external

to “F/R INDIC.” (flow direction)

voltage source not

and check (see Sect. 7.2)

supplying voltage.

with new status display:
Test ok, check previous
status display and external
voltage source, and replace
if necessary.
Test faulty, status output
defective, replace signal
converter (see Sect. 8.7) or
contact Krohne Service.

Output terminal B1 or B2

Change setting under

not set to “status output”

Fct. 3.07

Group D/I/P/S

Fault / Symptom

Caus

Remedial action

D / I / P / S 1

Unsteady display and

Electrical conductivity

Increase time constant under

outputs

of product too low,

Fct. 1.02.

time constant too low.

D / I / P / S 2

No display and no

Power OFF

Switch power on.

function of outputs

Check power fuse(s)

Replace if defective,

F1 (F1 + F2 for DC).

see Sect. 8.1.

Group C

Fault / Symptom

Cause

Remedial action

C 1

No function

Connection is incorrect

Connect properly,
see Sect. 2.6 + 2.7

Control input C

Check connection, cables

or external voltage

and external voltage source,

source defective

see Sect. 2.6 + 2.7

Output terminal

Change setting under

B1 or B2 not set to

Fct. 3.07.

“control input”