Part A System installation and start-up

1

Electrical connection: power supply

1.1

Location and important installation notes PLEASE NOTE !

Electrical connection in accordance with VDE 0100 ”Regulations governing heavy-current
installations with line voltages up to 1000 V” or equivalent national regulations.

Do not cross or loop cables inside the terminal compartment.

Use separate cable entries (PG screwed conduit entries) for power supply,
field current cables, signal lines, outputs and inputs.

Hazardous locations are subject to special regulations, see Section 6.1 and special
installation instructions for hazardous-duty (”Ex”) versions.

Protect flowmeters or switchgear cabinets with built-in devices from direct sunlight.
Fit a sunshade if necessary.

When installed in switchgear cabinets, signal converters must be adequately cooled,
e.g. use fans or heat exchangers.

Do not expose signal converters to intense vibration.

Only for separate systems/signal converters (F Versions)

Keep distance between primary head and signal converter as short as possible.
Refer to Sect. 1.3.4 for maximum permissible length of signal and field current cables.

Use the supplied Krohne signal cable A (Type DS), standard length 10 m (33 ft).

Always calibrate primary head and signal converter together. Therefore, when installing,
ensure primary constant GK is identical; refer to instrument nameplate for the primary
head. If the GK is not identical, set the signal converter to the GK of the primary head.
Refer also to Sections 4 and 8.5.

Dimensions of signal converter; refer to Section 10.2.