KROHNE IFC 090 Converter EN User Manual
Page 27
4/8
Fct.
Text
Description and settings
2.00
TEST
Test menu
2.01
TEST Q
Test measuring range Q
Precautionary query
• SURE NO
Press
↵
key to return to Fct. 2.01 “TEST Q”.
• SURE YES
Press
↵
key, then use
↑
key to
select value:
-110 / -100 / -50 / -10 / 0 / +10 / +50 / +100 / +110 PCT.
of set full-scale range Q
100%
.
Displayed value present at outputs I and P.
Press
↵
key to return to Fct. 2.01 “TEST Q”.
2.02
HARDW. INFO
Hardware information and error status
Before consulting factory, please note down all 6 codes.
→
MODUL ADC
X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press
↵
key to transfer to
“MODUL IO“.
→
MODUL IO
X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press
↵
key to transfer to
“MODUL DISP.“.
→
MODUL DISP.
X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press
↵
key to return to Fct. 2.02
“HARDW. INFO“.
3.00
INSTALL.
Installation menu
3.01
LANGUAGE
Select language for display texts
• GB / USA (English)
• F (French)
• D (German)
• others on request
Press
↵
key to return to Fct. 3.01
“LANGUAGE”.
3.02
FLOWMETER
Set data for primary head
→
DIAMETER
Select size from meter size table
• DN 10 - 1000 mm equivalent to 3/8 - 40 inch
Select with
↑
key.
Press
↵
key to transfer to subfunction “FULL SCALE ”.
→
FULL SCALE
Full-scale range for flow Q
100%
To set, refer to Fct. 1.01 “FULL SCALE” above.
Press
↵
key to transfer to subfunction “GK VALUE”.
→
VALUE P
Pulse value (Fct. 1.06
“VALUE P”
) has been changed.
With the “old” pulse values the output frequency (F)
would have been exceeded or not reached.
P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
Check new values!
→
GK VALUE
Set primary constant GK
see primary head nameplate.
Range:
• 1.0000 - 9.9999
Press
↵
key to transfer to subfunction
“FIELD. FREQ.”.
→
FIELD FREQ.
Magnetic field frequency
Values: 1/2, 1/6, 1/18 and 1/36 of power frequency, see nameplate.
Press
↵
key to transfer to subfunction “FLOW DIR.”;
DC units only: to transfer to subfunction
“LINE FREQ.”.
→
LINE FREQ.
Normal line frequency in your country
Please note: This function is only provided for units
with DC power supply to suppress line-frequency interference.
Values:
50 Hz
and
60 Hz
Press
↵
key to transfer to subfunction “FLOW DIR.”.
→
FLOW DIR.
Define flow direction (in F/R mode: forward flow).
Set according to direction of arrow on primary head:
• + DIR.
• – DIR.
Select using
↑
key.
Press
↵
key to return to Fct. 3.02 “FLOWMETER ”.