Installation and operating instructions IFC 020

10.3 IFC 020 signal converter

10/3

Versions

with display / control elements HART

®

and RS 485 interfaces

IFC 020 K

Compact version,converter directly mounted on the signal

converter version

IFC 020 F

Seperate housing fieldversion

IFC 020 E

Seperate housing, 19” Plug-in unit

Current output

Function

– all operating data settable, galvanically isolated

Current:

ranges

0 – 20 mA and 4 – 20 mA

Active mode

max. 500

load

Passive mode

external voltage:

15 ... 20 V DC

20 ... 32 V DC

load: min. ... max.

0 ... 500

250 ... 750

Error identification

0 / 3.6 / 22 mA and variable

Forward/reverse flow measurment

direction identified via status output

Pulse output

Function

– all operating data settable, galvanically isolated

– digital pulse division, interpulse period non-uniform,

therefore if frequency and cycle meters connected

allow for minimum counting interval:

gate time, totalizer

Pulse rate for Q=100%

10, 100 or 1000 pulses/sec.(=Hz), blocked or to your choice

per m

3

, Litre US Gallons or freely selectable unit.

Activ mode

connection: electronic totalizers

voltage: approx. 15 V DC, from current output

load: I

max

< 23 mA, operation without current output

load:

I

max

<

2

3 mA, operation with current output

Passive mode

connection: electronic or electromechanical totalizers

voltage: external, U

ext

32 V DC/

24 V AC

load: I

max

150 mA

Pulse width

– 50, 100, 200 or 500 msec.,selectable at frequencies < 10Hz

– symetric at frequencies < 10Hz

Forward/reverse flow measurement

flow direction identified via status output

Status output (passive)

Function

settable as indicator for flow direction, errors or trip point

Connection

voltage: external, U

ext

32 V DC/

24 V AC

load current: I

max

150 mA

Control input (passive)

IFC 020 E only

Function

– adjustable for totalizer reset, or reset setting

outputs to min. values

– initiate function by “low” or “high” control signals

Control signals

U

max

: 24 V AC

32 V DC (any polarity)

low:

1,4 V

2 V

high:

3 V

4 V

Time constant

0.2 – 99.9 s, adjustable in increments of 0.1 second

Low-flow cutoff

Cutoff “on” value: 1 – 19% of Q

100%

, adjustable

Cutoff “off” value: 2 – 20% in 1% increments

1000

P

100%

[Hz]

}