Fct.

Text

Description and settings

1.06

PULS B1

Pulse output P

FUNCT. P

Select function for pulse output P

• OFF (switched off)

• 1 DIR. (1 flow direction)

• 2 DIR. (forward/reverse flow, F/R flow measurement)

Press key

to transfer to subfunction “SELECT P”.

SELECT P

Select pulse type

• 100 HZ

• PULSE/VOL. (pulses per unit volume, flowrate)

• 1000 HZ

• PULSE/TIME (pulses per unit time for 100% flowrate)

Press key

to transfer to subfunction “PULSWIDTH”.

With selection 100 HZ and 1000 HZ,

return to Fct.1.06 PULSOUTPUT P (pulswidth 50% cyclic).

PULSWIDTH

Select pulse width

• 50 msec • 100 msec • 200 msec • 500 msec • 1 sec

Press key

to transfer to subfunction “VALUE P”.

VALUE P

Set pulse value per unit volume (appears only when

“PULSE/VOL.” set under “SELECT P” above)

• xxxx PulS/m3 • xxxx PulS/Liter • xxxx PulS/US.Gal

• xxxx PulS/ user-defined unit, factory-set is “Liter” or “US M.Gal” (see Fct. 3.05)

Setting range “xxxx” is dependent on the pulse width and the

full-scale range: P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Press key

to return to Fct.1.06 “PULS. OUTPUT P”.

VALUE P

Set pulse value per unit time (appears only when “PULSE/TIME”

set under “SELECT P” above)

• xxxx PulSe/Sec (=Hz) • xxxx PulSe/min • xxxx PulSe/hr

• xxxx PulSe/user-defined unit, factory-set is “hr” (see Fct. 3.05)

Setting range “xxxx” is dependent on the pulse width, see above

Press key

to return to Fct.1.06 “PULS. OUTPUT P””.

1.07

STATUS. S

Status output S

• ALL ERROR

• FATAL ERROR

• OFF

• ON

• F/R INDIC. (F/R indication for forward/reverse measurement)

• TRIP. POINT

Setting range: 002 - 115 PERCENT

Press

key to transfer to number setting

Press

key to return to Fct. 1.07 “STATUS. S“.

1.08

CONTROL E

Control input E (IFC 020 E only)

(IFC 020 E only)

• OFF (switched off)

• OUTP. ZERO (set outputs to “min.values”)

• TOTAL. RESET (reset totalizers)

• ERROR. RESET(delete error messages)

Press

key to return Fct. 1.08 “Control E”.

Installation and operating instructions IFC 020

Fct.

Text

Description and settings

2.00

TEST

Testmenu

2.01

TEST Q

Test measuring range Q

Precautionary query

• SURE NO

Press

key to return to Fct. 2.01 “TEST Q”.

• SURE YES

Press

key, then use

or

key to

select value: -110 / -100 / -50 / -10 / 0 / +10 / +50 / +100 / +110 PCT.

of set full-scale range Q

100%

.

Displayed value present at outputs I and P.

Press

key to return to Fct. 2.01 “TEST Q”.

2.02

HARDW. INFO

Hardware information and error status

Before consulting factory, please note down all 6 codes.

MODUL ADC

X . X X X X X . X X

Y Y Y Y Y Y Y Y Y Y

Press

key to transfer to “MODUL IO“.

MODUL IO

X . X X X X X . X X

Y Y Y Y Y Y Y Y Y Y

Press

key to transfer to “MODUL DISP.“.

MODUL DISP.

X . X X X X X . X X

Press

key to transfer to “MODUL RS

Y Y Y Y Y Y Y Y Y Y

MODUL RS

X . X X X X X . X X

Press

key to return to Fct. 2.02

Y Y Y Y Y Y Y Y Y Y

“HARDW. INFO“.

4/7