Installation and operating instructions IFC 020

4/5

π

4

π

4

Q

actual flowrate

Q

100%

100% flow = full scale range

Q

max

=

DN

2

x v

max

/ max. full-scale range (Q

100%

)

at v

max

= 12 m/s / 40 ft/s

Q

min

=

DN

2

x v

min

/ min. full-scale range (Q

100%

)

at v

min

= 0.3 m/s / 1 ft/s

S

Status output

SMU

Low-flow cutoff for I and P

v

Flow velocity

v

max

Max. flow velocity (12 m/s / 40 ft/s) at Q

100%

v

min

Min. flow velocity (0.3 m/s / 1 ft/s) at Q

100%

F/R

Forward/Reverse flow at F/R operation

π

4

4.4 Table of settable functions

Abbreviations used

C

Control input (IFC 020 E only)

DN

Nominal size, meter size

F

max

Highest frequency of pulse output

F

min

Lowest frequency of pulse output

F

M

Conversion factor volume for any unit,

see Fct. “FACT. VOL.”

F

T

Conversion factor time for any unit,

see Fct. 3.05 “FACT. Time”

GK

Primary constant

I

Current output

I

0%

Current at 0% flow

I

100%

Current at 100% flow

P

Pulse output

P

max

= F

max

/ Q

100%

P

min

= F

min

/ Q

100%

π

4

Fct.

Text

Description and settings

1.00

OPERATION

Operations menu

1.01

FULL SCALE

Full-scale range for flowrate Q

100%

Select unit

• m

3

/hr

• Liter/Sec

• US.Gal/min

• user unit, factory set is “Liter/hr” or “US MGal/day” (see Fct. 3.05)

Press

key to transfer to number setting.

Setting ranges

The ranges are dependent on the meter size (DN) and the flow

velocity (v): Q

min

= DN

2

x v

min

Q

max

= DN

2

x v

max

Nom. dia./meter size

v

min

= 0,3 m/s (1 ft/s) v

max

= 12 m/s (40 ft/s)

• DN 2.5–1000 /

1

/

10

”–40”: 0.0053 –

0

33 900 m

3

/hr

0.0237 –

152 000 US.Gal/min

Press

key to return to Fct. 1.01 FULL SCALE.

VALUE P

Pulse value (Fct. 1.06 “VALUE P”) has been changed.

With the “old” pulse values the output frequency (F)

would have been exceeded or not reached.

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Check new values!

1.02

TIMECONST.

Time constant

Select:

• ALL (applies to display and all outputs)

• ONLY I+S (only display, current and status outputs)

Press

key to transfer to number setting.

Range:

• 0.2 – 99.9 Sec

Press

key to return to Fct. 1.02 TIMECONST.

1.03

L.F.CUTOFF

Low-flow cutoff (SMU)

• OFF (fixed values: ON = 0.1% / OFF = 0.2%

at 100Hz and 1000Hz, see Fct. 106, 1% resp. 2%)

• PERCENT (variable values)

ON OFF

1 – 19% 2 – 20%

Press

key to transfer to number setting.

Note: Cutoff “off” value must be greater than cutoff “on” value.

Press

key to return to Fct. 1.03 L.F. CUTOFF.