KROHNE IFC 020 Converter EN User Manual
Page 43
Installation and operating instructions IFC 020
5/12
Applications 5.15
Measuring point identification 5.16
Fct. 3.06 APPLICAT.
Press key
→
.
→
FLOW = Set characterization for the flow,
press key
• STEADY
(flow is steady)
• PULSATING
(pulsating flow, e.g. due to reciprocating pumps,
see also Sect. 6.4, 6.5 and 6.6
“Special-case applications”)
Press
↵
key to return to Fct. 3.06 APPLICAT.
}
select function using
key
↑
or
↓
Fct. 3.07 MEAS.PNT.
Press key
→
.
Set measuring point identification using max. 10 characters , (e.g. TQ1_532197)
You can use any out of A...Z / a...z / 0...9 or _ (underscore)
- Select character by pressing key
↑
or
↓.
- Use key
→
to shift one place to the left, back with
←
key
.
Press key
↵
to return to Fct. 3.07 MEAS.PNT.
Communication interface 5.17
Control input E (IFC 020 E only) 5.18
Fct. 3.08 COM
Press key
→
.
Specify function
• OFF
(Switched off)
• HART
(HART Interface)
• KROHNE (RS 485 Interface)
Use key
↑
or
↓
to set ADDRESS 00-15 after selecting "HART" with
→
key.
Press key
↵
dto return to Fct. 3.08 COM.
Use key
↑
or
↓
to set ADDRESS 000-239 after selecting "KROHNE" (RS 485). Press key
→
and
set up BAUD RATE: • 1200 • 2400 • 4800 • 9600 • 19200. Use key
↑
or
↓
to select baud rate.
Press key
↵
to return to Fct. 3.08 COM.
}
select function using key
↑
or
↓
Fct. 1.08 CTRL.INP. E
Press key
→
.
Select control input function, using key
↑
or
↓
.
• OFF
(Switched off, out of function)
• ZERO OUTPUT
(Set output and display to "Min. values")
• RESET COUNT.
(Reset counter)
• RESET ERROR
(Delete / confirm error message)
Press key
↵
to return to Fct. 1.08 CTRL.INP. E
For factory-set preferences refer to the start-up protocol and Section 3.2.