Installation and operating instructions IFC 020

Current output

same error limits as above, additionally ± 10

µ

A

Reproducibility and

repeatability

0.1% of MV, minimum 1 mm/s / 0.04 inch/s at constant flow

External influences

typical values

maximum values

Ambient temperature

Pulse output

<

0.003% of MV (1)

0.01

5

% of MV (1)

Current output

<

0.01

3

% of MV (1)

0.025% of MV (1)1

Power supply

< 0.02

3

% of MV

0.05

5

% of MV

at 10% variation

Load

< 0.01

3

% of MV

0.02

5

% of MV

at max. permissible load,

see pages 6 and 7

(1) All Krohne signal converters undergo burn-in tests, duration minimum 20 hours at varying ambient temperatures

– 20 to + 60°C/– 4 to + 140°F. The tests are controlled by computers.

Error limits at reference conditions 10.2

Display, digital values, pulse output

F maximum error in % of measured value (not typical values)

v Flow velocity in m/s and ft/s

Reference conditions similar to EN 29104

Product

water at 10 – 30°C/50 – 86°F

Electrical conductivity

> 300

µ

S/cm

Power supply (rated voltage)

U

N

(± 2%)

Ambient temperature

20 – 22°C/68 – 71.6°F

Warm-up time

60 min

Max. calibration equipment error 10x smaller than F

Inlet/outlet runs

10

x

DN/2

x

DN (DN = meter size)

Primary head

properly grounded and centered

Calibrated on EN 17025 accredited calibration equipment in direct volumetric comparsion.

10/2

Type/Meter size

Maximum error in % of measured value (MV) ...

Curve

DN mm

DN 2.5 – 6

DN 2.5 – 6(1)

DN 10

DN 10

inch

1

/

10

” –

1

/

4

DN 2.5 – 6(1)

3

/

8

DN 10

v

1 m/s /

3 ft/s

± 0.5% of MV

DN 2.5 – 6(1)

± 0.3% of MV

DN 10

v

<

1 m/s /

<

3 ft/s

± (0.4% of MV + 1 mm/s)

± (0.4% of MV + 0.04 inch/s

± (0.2% of MV + 1 mm/s)

± (0.2% of MV + 0.04 inch/s)

B

D

A

E

per 1 K / 1.8°F temperature

variation

}

* IFS 6000 F (DN 2.5-4 and 1/10”-1/6”) additional error ±0.3% of MV