beautypg.com

KROHNE IFC 020 Converter EN User Manual

Page 60

background image

Installation and operating instructions IFC 020

8/3

8.2 Replacement of power supply fuse

Switch off power supply before opening the housing!

Power PCB

Power supply Fuse F1

IFC 020 K and IFC 020 F for Section 8.3

Location and position of voltage selector

Rating

Order No.

1. AC version

230/240 V AC

125 mA T

5.06627

115/117 V AC

200 mA T

5.05678

2. AC version

200 V AC

125 mA T

5.06627

100 V AC

200 mA T

5.05678

3. AC version

48 V AC

400 mA T

5.05892

24 V AC

800 mA T

5.08085

DC version

24.V DC

F 1

F 1

F 1

F 1

IFC 020 K and IFC 020 F

IFC 020 E

Refer to Sect. 8.1 for Figs. A and B.
1)

Unscrew the 4 recessed head screws

(Fig. A), and remove transparent cover.

2)

Unscrew the recessed head screw

(Fig. B) and remove the black plastic

cover.

3)

Take out old and insert new power fuse

F1, on the left next to the green

connecting terminals.

Please refer to the following table for

fuse rating and order number.

4)

Reassemble in reverse order,

points 2) – 1) above.

Refer to Sect. 8.1 for Figs. 1 and 2.
1)

Unscrew the 4 recessed head

screws (S1), on the front (Fig. 1).

2)

Carefully remove the plug-in unit from

the subassembly support.

3)

Replace power fuse F1 back of

cassette(Fig. 2). Refer to the following

table for fuse rating and order number.

4)

Reassemble in reverse order,

points 2) - 1) above.

IFC 020 E only

IFC 020 E only