Installation and operating instructions IFC 020

4/10

Error messages in measuring mode 4.5

The following list gives all errors that can occur during process flow measurement.

Errors shown in display when “YES“ set in Fct. 1.04 DISPLAY, subfunction “DISP. MSG.“.

Error messages

Description of error

Error clearance

LINE INT.

Power failure Note:

Cancel error in RESET/QUIT. menu

no counting during power failure

Reset totalizer if necessary.

CUR. OUTP. I

Current output overranged.

Check and if necessary correct

(Flow > measuring rage)

instrument parameters. After

elimination of cause, error message

deleted automatically.

PULSOUTP. P

Pulse output overranged.

Check and if necessary correct

Note:

instrument parameters. After

totalizer deviation possible.

elimination of cause, error message

deleted automatically.

TOTALIZER

Totalizer has been reset

Cancel error message in

RESET/QUIT. menu, see Section 4.6.

ADC

Analog / digital

Error message deleted automatically

converter overranged

after elimination of cause.

FATAL. ERROR

Fatal error, all outputs set

Please consult factory.

to “min. values“