Fault / Symptom

Cause

Remedial action

P 1

Totalizer connected but does

Incorrect connection/polarity

Connect properly,

not count any pulses

see Sect. 2.4.

Totalizer or external voltage

Check output (see Sect. 7.2)

source defective.

with new totalizer:

Test ok, check connection

cables and previous

totalizers and external voltage

source, and replace if

necessary.

Test faulty, pulse output

defective, replace signal

converter (see Sect. 8.4)

or contact Krohne Service.

Current output is external

Check connection and cables,

voltage source, short circuit

see Sect. 2.4. Voltage between

or current / pulse output

I+ and I

approx. 15 V.

defective

Switch off device, eliminate

short circuit, switch device

on again. If no function,

then current or pulse

output defective. Replace

signal converter (see Sect.

8.4) or contact Krohne

Service.

Pulse output is deactivated,

Switch on under Fct. 1.06.

see Fct. 1.06.

P 2

Unsteady pulse rate

– Process product

Increase or switch on

conductivity too low,

time constant, under Fct. 1.02

particles/air inclusions

orswitch on if necessary.

too large or inhomogeneous

– Pulsating flow

– Time constant too low

or switched off

P 3

Pulse rate too high or

Incorrect setting for

Change setting under Fct. 1.06.

too low.

pulse output.

Group S

Fault / Symptom

Cause

Remedial action

S 1

No function

Incorrect connection/polarity

Connect properly,

of status display

see Sect. 2.4.

Status display or output

Set status output under Fct. 1.07

defective or external

to “F/R INDIC.“ (flow direction)

voltage source not

and check (see Sect. 7.2)

supplying voltage.

with new status display:

Test ok, check previous

status display and external

voltage source, and replace

if necessary.

Test faulty, status output

defective, replace signal

converter (see Sect. 8.4) or

contact Krohne Service.

Group D/I/P/S

Fault / Symptom

Cause

Remedial action

D / I / P / S 1

Unsteady display and

– Process product

Increase time constant,

outputs

conductivity too low,

under Fct. 1.02

particles/air inclusions

too large or inhomogeneous

– Pulsating flow

– Time constant too low

D / I / P / S 2

No display and no

Power OFF

Switch power on.

function of outputs

Check power fuse(s)

Replace if defective,

F1 (F1 and F2 for DC).

see Sect. 8.2.