5.11 Primary head

Please refer to Sect. 3.2 “factory settings”

Fct. 3.02 FLOW METER

Press

key.

DIAMETER = set meter size (see instrument nameplate) press

key

Select size from table of meter sizes:

• DN 2.5 - 1000 mm equivalent to

1

/

10

- 40 inch

Select using

or

key.

Transfer to subfunction “FULL SCALE” with

key.

FULL SCALE = set full-scale range, press

key.

Set as described in Sect. 5.1.

Transfer to subfunction “GK VALUE” with

key.

Note: if “VALUE P” is displayed after pressing

key.

PULSE/VOL. is set under Fct. 1.06 PULS.OUTP. P, subfunction “SELECT P”. Because the

full-scale range Q

100%

has been changed, the output frequency (F) of the pulse output is over- or

undershot:

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Change pulse value accordingly, see Sect. 5.7 pulse output P, Fct. 1.06.

GK VALUE = set primary constant GK, press

key.

• 1.0000 - 9.9999 (note information on instrument nameplate, do not change setting !)

Change flashing digit (cursor) with

or

key.

Shift cursor 1 place to right or left with

or

key.

Transfer to subfunction “FIELD FREQ.” with

key.

FIELD FREQ. = set magnetic field frequency, press

key

• 1/2 • 1/18 (1/2, 1/6 1/18 or 1/36 of power frequency,

• 1/6 • 1/36 see instrument nameplate, do not change setting)

Select using

or

key.

Transfer to subfunction “FLOW DIR.” with

key.

(only for units with DC power supply, transfer to subfunction “LINE FREQ”).

LINE FREQ. = normal line frequency in your country, press

key.

(Please note, applicable for flowmeters equipped with DC in power supply!)

• 50 Hz

Select using the

or

key.

• 60 Hz

Transfer to subfunction “FLOW DIR.” with

key.

FLOW DIR. = set flow direction, press

key.

• + DIR.

(for identification of flow direction, see “+” arrow on primary head;

• - DIR.

for F/R mode, identifies the “positive” flow direction)

Select using the

or

key.

Press

key to return to Fct. 3.02 FLOW METER.

Zero check, see Fct. 3.03 and Sect. 7.1.

}