Group D

Display shows . . .

Cause

Remedial action

D 1

LINE INT.

Power failure.

Delete error message in

Note: no counting during

RESET/QUIT. menu. Reset

power failure.

totalizer(s) if need be.

D 2

CUR.OUTP. I

Current output overranged.

Check instrument parameters

and correct if necessary.

Error message deleted

automatically after cause

has been eliminated.

D 3

PULS.OUTP. P

Pulse output overranged.

Check instrument parameters,

Note: totalizer deviation

correct if necessary, and reset

possible

totalizer(s). Error message

deleted automatically

after cause has been

eliminated.

D 4

ADC

Analog/digital converter

Error message deleted

overranged.

automatically after cause

has been eliminated.

D 5

FATAL. ERROR

Fatal Error,

Replace signal converter

all outputs set to “min.”

(see Sect. 8.4) or contact

values.

Krohne Service, having

first noted down hardware

information and error status,

see Fct. 2.02.

D 6

TOTALIZER

Counts lost

Delete error message in

(overflow, data error)

RESET/QUIT. menu.

D 7

„STARTUP“

Hardware fault,

Replace signal converter

cyclic flashing

Watchdog activated.

(see Sect. 8.4) or contact

Krohne Service.

D 8

BUSY

Displays for flow,

Change setting in

totalizers and errors disabled.

Fct. 1.04.

D 9

Unsteady display

– Process product

Increase time constant,

conductivity too low,

see Fct. 1.02, or switch on.

particles/air inclusions

too large or inhomogeneous

– Pulsating flow

– Time constant too low or

switched off.

D 10

No display

Power OFF.

Switch on power

Check power fuse(s)

Replace if defective

F1 (F1 and F2 with DC).

(see Sect. 8.2).

Group I

Fault / Symptom

Cause

Remedial action

I 1

Receiver instrument

Incorrect connection/polarity

Connect properly,

indicates “0”.

see Sect. 2.4.

Receiver instrument or current

Check output (see Sect. 7.2)

output defective.

with new milliammeter:

Test ok, check connection

cables and receiver instrument,

replace if necessary.

Test faulty, current output

defective. Replace signal

converter (see Sect. 8.4)

or contact Krohne Service

Current output disabled,

Activate under Fct. 1.05.

see Fct. 1.05
Short-circuit between current

Check connection and cables,

output and pulse output.

see Sect. 2.4. Voltage between

I+ and I

approx. 15 V.

Switch off device, eliminate

short-circuit, and switch

device on again.

I 2

Unsteady display

– Process product

Increase time constant,

conductivity too low,

see Fct. 1.02

particles/air inclusions

too large or inhomogeneous

– Pulsating flow

– Time constant too low or

switched off.

Installation and operating instructions IFC 020