beautypg.com

Technical data, 5 dynamic output behaviour – KROHNE OPTIBAR DP 7060 C EN User Manual

Page 71

background image

TECHNICAL DATA

7

71

OPTIBAR DP 7060 C

www.krohne.com

04/2014 - 4003292001 - MA OPTIBAR DP 7060 C R01 en

7.5 Dynamic output behaviour

These parameters depend on the filling medium, temperature and, if applicable, the chemical
seal.

Figure 7-2: Behaviour for an sudden change of the process variable.
t

1

- dead time; t

2

- rise time; t

3

- step response time

1 Process variable
2 Output signal

Dead time (t1)
[ms]

T90% (t2)
[ms]

Step response time
(t3)
[ms] 1

30 mbar / 0.44 psi

90

115

205

100 mbar / 1.50 psi

60

95

155

500 mbar / 7.3 psi

75

135

3 bar / 43.51 psi

60

120

16 bar / 232.1 psi

1 Step response time is the sum of dead time and T90%

.book Page 71 Friday, May 2, 2014 10:25 AM