Start-up – KROHNE OPTIBAR DP 7060 C EN User Manual
Page 48
5
START-UP
48
OPTIBAR DP 7060 C
www.krohne.com
04/2014 - 4003292001 - MA OPTIBAR DP 7060 C R01 en
5.4.2 Display
Menu language In this menu item you can set the desired language.
Factory settings: English
German, English, French, Spanish, Portuguese, Italian, Dutch, Russian, Turkish, Polish and Czech
Displayed value
1 and 2
In this menu item you can define how the measured value should be presented on the display.
The factory setting is "Linear percent".
"Level"
Filling height or Pressure (Adjustment units),
Static pressure, Percent, Scaled, Current output,
Linear percent, Meas. cell temp. and Electronics
temperature
"Differential pressure"
Differential pressure, Static pressure, Percent,
Scaled, Current output, Linear percent, Meas.
cell temp. and Electronics temperature
"Flow"
Flow, Differential pressure, Static pressure,
Pressure/percent, Scaled, Current output, Linear
percent, Totalizer 1, Totalizer 2, Meas. cell temp.
and Electronics temperature
"Interface"
Interface height, Differential pressure, Percent,
Scaled, Current output, Linear percent, Meas.
cell temp. and Electronics temperature
"Density"
Density, Differential pressure, Percent, Scaled,
Current output, Linear percent, Meas. cell temp.
and Electronics temperature
Backlight
A backlight on the display is available, which can be turned on or off in this menu. By default, this
function is disabled.
Off, On
.book Page 48 Friday, May 2, 2014 10:25 AM