7
TECHNICAL DATA
68
OPTIBAR DP 7060 C
www.krohne.com
04/2014 - 4003292001 - MA OPTIBAR DP 7060 C R01 en
Electrical connections
Mechanical
Mechanical
Mechanical
Mechanical
Cable gland
M20 x 1.5 mm
Cable diameter
5...9 mm / 0.20...0.35"
6...12 mm / 0.24...0.47"
10...14 mm / 0.39...0.55"
Cable entry
1/2 NPT
Blind plug: M20 x 1.5 mm, 1/2-14 NPT
Closing cap: M20 x 1.5 mm, 1/2-14 NPT
Connector option: M12 x 1, Harting HAN 7D,8D, 7/8 FF
Core cross-section
Solid wire with cords: 0.2 mm...2.5 mm
2
/ AWG 24...14
Cord with wire end sleeve: 0.2 mm...1.5 mm
2
/ AWG 24...16
Electrical
Electrical
Electrical
Electrical
Supply voltage
Non-Ex device: 9.6...35 VDC
Ex ia device: 9.6...30 VDC
Ex d device: 9.6...35 VDC
Ex ia d device: 15...35 VDC
Reverse polarity
protection
Integrated
Permissible residual ripple
Non-Ex devices
for U
n
12 VDC (9.6 < UB < 14 VDC) ≤0.7 V
eff
(16...400 Hz)
Ex ia devices,
Ex ia d devices
for U
n
24 VDC (18 < UB < 35 VDC) ≤1.0 V
eff
(16...400 Hz)
Load
R
L,max
=(UB-9.6) / 22 mA
Potential connection
in device
Electronics: not electrically isolated
Ground terminal: galvanically connected with process connection
Overvoltage category
III
Protection class
II
Inputs and outputs
Output signal
Output signal
Output signal
Output signal
Output signal
4...20 mA / HART
®
version 7.3
3.8...20.5 mA (factory setting acc. to. NAMUR recommendation)
Signal resolution
0.3 µA
Error signal of current
output (adjustable)
High alarm ≥21 mA
Low alarm ≤3.6 mA, last valid measurement
Max. output current
21.5 mA
Boot-up current
≤10 mA for 5 ms after switching on, then ≤3.6 mA
Damping
0...999 seconds, adjustable
.book Page 68 Friday, May 2, 2014 10:25 AM