beautypg.com

Start-up – KROHNE OPTIBAR DP 7060 C EN User Manual

Page 57

background image

START-UP

5

57

OPTIBAR DP 7060 C

www.krohne.com

04/2014 - 4003292001 - MA OPTIBAR DP 7060 C R01 en

5.8.2 Check 4...20 mA signal

Connect a multimeter in the suitable measuring range according to the wiring plan.

5.8.3 Error messages via the display and operating module

Depending on the reason for the fault and the measures taken, the steps described previously
may need to be carried out again.

5.8.4 Change electronic insert

In case of a defect, the electronic insert can be exchanged by the user against an identical type. If
no electronic insert is available on site, it can be ordered from the respective local sales
representative. To order a replacement, the serial number is required. This is located on the
nameplate of the device or on the delivery note.

Error code

Cause

Action or description

4...20 mA signal is missing

Faulty connection to power

supply

Check connection and if

necessary correct according to

wiring plan

No power supply

Check cable for breaks; repair if

necessary

Operating voltage too low or

load-resistance too high

Check, adapt if necessary

Signal is >22 mA or <3.6 mA

Electronic module or sensor

defective

Exchange the device or send

device for repair

DANGER!
In hazardous area applications, the regulations for the wiring of intrinsically safe circuits must
be observed.

Error code

Cause

Action or description

E013

No measurement value available

or pressure greater than nominal

range

Exchange the device or send

device for repair

E017

Adjustment span too small

Repeat with modified values

E036

No executable signal converter

software

Carry out software update or

send device for repair

E041

Hardware error

Exchange the device or send

device for repair

WARNING!
Installation, assembly, start-up and maintenance may only be performed by personnel trained in
explosion protection

explosion protection

explosion protection

explosion protection. Additional regional standards, safety directives and laws must be observed
at all times.

.book Page 57 Friday, May 2, 2014 10:25 AM